Análise da resistência de fios cirúrgicos utilizados no reparo do manguito rotador / Strength analysis of surgical sutures used to repair the rotator cuff
Guimarães, Marcus Valladares; Carvalho Júnior, Lúcio Honório de.
  • Guimarães, Marcus Valladares; Universidade Federal de Minas Gerais. Belo Horizonte. BR
  • Carvalho Júnior, Lúcio Honório de; ociedades Brasileiras de Cirurgia do Joelho, Artroscopia e Cirurgia do Ombro e Cotovelo.
Rev. bras. ortop ; 43(7): 287-292, jul. 2008. ilus, tab
Artículo en Portugués | LILACS | ID: lil-492431
RESUMO

OBJETIVO:

Analisar a resistência final suportada por dois tipos de fio e pela interface âncora-fio em três diferentes implantes tipo âncora submetidos à tração contínua, com o propósito de indicar o mais apropriado para sutura de estruturas como o manguito rotador do ombro.

MÉTODOS:

Os testes foram feitos em três grupos grupo 1 - parafuso âncora de titânio com fio Ethibond® #2; grupo 2 - parafuso âncora de titânio com fio Fiber Wire® #2; e grupo 3 - parafuso âncora de titânio 5,0mm Corkscrew®.

RESULTADOS:

No grupo 1, a carga máxima média até a soltura foi de 124,5 newtons (N), sendo a mínima de 105N e a máxima de 180N, com variação de 75N entre os dois extremos e desvio-padrão de 23,03N. No grupo 2, a carga máxima média até a soltura foi de 298N, mínima de 230N, máxima de 375N, com variação de 145N e desvio-padrão de 44,73N. No grupo 3, foi obtida a maior carga máxima média, sendo esta de 272N, a mínima de 205N e a máxima de 340N, com variação de 135N. A análise estatística demonstrou diferença significativa entre as médias de cargas resistidas entre os grupos 1 x 2 e 1 x 3 (p < 0,001).

CONCLUSÃO:

Observou-se semelhança na resistência final obtida com o fio Fiber Wire® independentemente da âncora usada. O fio Ethibond #2 apresentou resistência inferior, menor que a desejada para a sutura de estruturas como do manguito rotador do ombro.
ABSTRACT

OBJECTIVE:

To analyze the final strength withstood by two types of suture threads and by the anchor-suture interface in three different anchor implants submitted to continuous traction for the purpose of choosing the most appropriate to suture structures such as the shoulder rotator cuff.

METHOD:

Tests were performed in three groups Group 1 - titanium anchor screw with Ethibond® #2 suture, Group 2 - titanium anchor screw with Fiber Wire® #2 suture, and Group 3 - Corkscrew® 5.0 mm anchor screw.

RESULTS:

In Group 1, mean maximum load till loosening was 124.5 124,5 Newtons (N), the minimum load was 105 N, and the maximum load was 180 N, with a range of 75 N between the two extremes, and standard deviation of 23.03 N. In Group 2, the mean maximum load till loosening was 298 N, the minimum load was 230 N, the maximum load was 375 N, with a range of 145 N, and standard deviation of 44.73 N. In Group 3, the highest mean maximum load was measured, of 272 N, the minimum load was 205 N, and the maximum load of 340 N, with a range of 135 N. Statistical analysis showed a significant difference between the mean loads withstood between groups 1 and 2, and between groups 1 and 3, (p < 0.001).

CONCLUSION:

Final strength was similar with Fiber Wire® suture irrespective of the anchor used. Ethibond #2 suture showed lower strength, lower than that desired to suture structures such as the shoulder rotator cuff.
Asunto(s)

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Hilos Ortopédicos / Técnicas de Sutura / Manguito de los Rotadores Límite: Humanos Idioma: Portugués Revista: Rev. bras. ortop Asunto de la revista: Ortopedia Año: 2008 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Universidade Federal de Minas Gerais/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Hilos Ortopédicos / Técnicas de Sutura / Manguito de los Rotadores Límite: Humanos Idioma: Portugués Revista: Rev. bras. ortop Asunto de la revista: Ortopedia Año: 2008 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Universidade Federal de Minas Gerais/BR