Your browser doesn't support javascript.
loading
Radial dilation of nephrostomy balloons: a comparative analysis
Hendlin, Kari; Monga, Manoj.
  • Hendlin, Kari; University of Minnesota. VAHCS Minneapolis. Departments of Urologic Surgery. Minneapolis. US
  • Monga, Manoj; University of Minnesota. VAHCS Minneapolis. Departments of Urologic Surgery. Minneapolis. US
Int. braz. j. urol ; 34(5): 546-554, Sept.-Oct. 2008. ilus, graf, tab
Artículo en Inglés | LILACS | ID: lil-500389
ABSTRACT

PURPOSE:

The dynamics of percutaneous balloon expansion may differ with increasing extrinsic compressive forces and increasing inflation pressures. This study compares the ability of percutaneous balloons to expand under different radial constrictive forces. MATERIALS AND

METHODS:

Three 30F nephrostomy balloons were tested Bard X-Force™, Boston Scientific Microvasive Amplatz Tractmaster™, and Cook Ultraxx™. With a super stiff guidewire in place, the balloon tip was secured by elevated vice grips on either side of the balloon. A string was wrapped around the balloon center once, and incremental increases in load were added (2g, 42g, 82g, and 122g) to represent increasing extrinsic compression. The balloon was inflated with a contrast agent and circumference changes were measured at increments of 4 ATM, 10 ATM, and burst pressure. Balloons were tested in triplicate for each load.

RESULTS:

All balloons were unable to reach 90 percent of their expected diameter with larger constrictive loads (122g) at low (4 ATM) and nominal (10 ATM) inflation pressures. Only the Bard and Cook balloons reached at least 90 percent of the expected diameter with a coefficient of variance (CV) less than 10 percent at burst pressure under the larger constrictive load (122g), 94.3 percent ± 6.7 percent, CV 7.1 percent and 96.3 percent ± 2.9 percent, CV 3.0 percent respectively. All balloons performed well under low constriction forces and reached at least 80 percent of the expected diameter by 10 ATM under all constrictive loads.

CONCLUSIONS:

The Bard X-Force and Cook Ultraxx percutaneous nephrostomy balloons achieved the most reliable radial dilation against large constrictive forces simulating fascial or retroperitoneal scar tissue.
Asunto(s)

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Ensayo de Materiales / Nefrostomía Percutánea Límite: Humanos Idioma: Inglés Revista: Int. braz. j. urol Asunto de la revista: Urología Año: 2008 Tipo del documento: Artículo País de afiliación: Estados Unidos Institución/País de afiliación: University of Minnesota/US

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Ensayo de Materiales / Nefrostomía Percutánea Límite: Humanos Idioma: Inglés Revista: Int. braz. j. urol Asunto de la revista: Urología Año: 2008 Tipo del documento: Artículo País de afiliación: Estados Unidos Institución/País de afiliación: University of Minnesota/US