Estudo comparativo da qualidade de imagem dos modos de aquisição da PET: validação de um protocolo para reduzir a dose de radiação / Comparative study on imaging quality in PET acquisition modes: validation of a protocol for reducing the radiation dose
Nogueira, Solange Amorim; Lederman, Henrique Manoel; Wagner, Jairo; Yamaga, Lilian Yuri; Cunha, Marcelo Livorsi da; Funari, Marcelo Buarque de Gusmão.
  • Nogueira, Solange Amorim; Hospital Albert Einstein. Setor de Medicina Nuclear. São Paulo. BR
  • Lederman, Henrique Manoel; Universidade Federal de São Paulo. Escola Paulista de Medicina. Departamento de Diagnóstico por Imagem. Disciplina de Diagnósticopor Imagem em Pediatria. São Paulo. BR
  • Wagner, Jairo; Hospital Israelita Albert Einstein. Setor de Medicina Nuclear. São Paulo. BR
  • Yamaga, Lilian Yuri; Hospital Israelita Albert Einstein. Setor de Medicina Nuclear. São Paulo. BR
  • Cunha, Marcelo Livorsi da; Hospital Israelita Albert Einstein. Setor de Medicina Nuclear. São Paulo. BR
  • Funari, Marcelo Buarque de Gusmão; Hospital Israelita Albert Einstein. Departamento de Imagem. São Paulo. BR
Radiol. bras ; 42(2): 103-107, mar.-abr. 2009. ilus, graf, tab
Artículo en Portugués | LILACS | ID: lil-513151
RESUMO
OBJETIVO: O presente experimento visa a validar um protocolo de aquisição em 3D na tomografia por emissão de pósitrons, em substituição ao modo 2D, de forma a reduzir a dose de radiação nos pacientes, sem perda da qualidade de imagens. MATERIAIS E MÉTODOS: Foram realizadas 27 simulações em equipamento Discovery ST, nos modos 2D com quatro minutos de aquisição e 3D com dois e quatro minutos. Utilizou-se um simulador do protocolo da National Electrical Manufacturers Association. No interior deste simulador estão inseridas seis esferas com diferentes diâmetros para a determinação da qualidade de imagem. As aquisições foram comparadas por três médicos nucleares, sem que eles identificassem o modo de aquisição. Cada observador atribuiu o valor igual a 1 quando alguma esfera não foi identificada ou valor 2 para esferas visíveis. RESULTADOS: A análise qualitativa pelo kappa generalizado demonstrou que a frequência de esferasvisíveis foi maior no modo 3D com quatro minutos (85%) e a porcentagem de concordância também foi maior (88,9%), com kappa generalizado = 0,725 [0,507;0,942]. CONCLUSÃO: O modo 3D com quatro minutos de aquisição e com menores atividades de FDG-18F pode ser utilizado em pacientes com biótipo equivalente ao simulador, sem perda de qualidade de imagem.
ABSTRACT
OBJECTIVE: The present study is aimed at validating a 3D acquisition protocol for positron emission tomography as a replacement for the 2D mode, to reduce the radiation dose delivered to patients, without any loss in the quality of images. MATERIALS AND METHODS: The study comprised 27 simulations in a Discovery ST equipment with four-minute 2D acquisitions, and two-minute and four-minute 3D acquisitions, utilizinga chest phantom according to the National Electrical Manufacturers Association protocol. Six spheres withdifferent diameters were inserted into this phantom as a means for determining the images quality. The images were blindly reviewed by three experienced nuclear physicians who did not know the acquisition modes. Each observer attributed a score 1 when one of the spheres was not identified, or 2 for visible spheres. RESULTS: The qualitative analysis based on generalized kappa coefficient demonstrated that the frequency of visible spheres was higher with four-minute 3D acquisitions (85%), with an also higher interobserver agreement (88.9%), generalized kappa = 0.725 [0.507;0.942]. CONCLUSION: The protocol with four-minute 3D acquisition with lower 18F-FDG activity can be utilized for patients with a biotype similar to the phantom, without any loss in the imaging quality.
Asunto(s)

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Imagenología Tridimensional / Tomografía de Emisión de Positrones Tipo de estudio: Estudio pronóstico / Investigación cualitativa Idioma: Portugués Revista: Radiol. bras Asunto de la revista: Diagnóstico por Imagen / Radiología Año: 2009 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Hospital Albert Einstein/BR / Hospital Israelita Albert Einstein/BR / Universidade Federal de São Paulo/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Imagenología Tridimensional / Tomografía de Emisión de Positrones Tipo de estudio: Estudio pronóstico / Investigación cualitativa Idioma: Portugués Revista: Radiol. bras Asunto de la revista: Diagnóstico por Imagen / Radiología Año: 2009 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Hospital Albert Einstein/BR / Hospital Israelita Albert Einstein/BR / Universidade Federal de São Paulo/BR