Your browser doesn't support javascript.
loading
Influência do manuseio do balão autoinflável neonatal sobre o pico de pressão e o volume corrente / Peak pressure and tidal volume are affected by how the neonatal self-inflating bag is handled
Bassani, Mariana A; Mezzacappa Filho, Francisco; Coppo, Maria Regina C; Marba, Sérgio T. M.
  • Bassani, Mariana A; Universidade Estadual de Campinas. Centro de Investigação em Pediatria. Campinas. BR
  • Mezzacappa Filho, Francisco; UNICAMP. CAISM. Departamento de Pediatria. Campinas. BR
  • Coppo, Maria Regina C; Universidade Estadual de Campinas. Centro de Investigação em Pediatria. Campinas. BR
  • Marba, Sérgio T. M; UNICAMP. Faculdade de Ciências Médicas. Departamento de Pediatria. Campinas. BR
J. pediatr. (Rio J.) ; 85(3): 217-222, maio-jun. 2009. tab
Artículo en Inglés, Portugués | LILACS | ID: lil-517869
RESUMO
OBJETIVO: Avaliar a influência do manuseio do ventilador pulmonar manual neonatal com bolsa autoinflável sobre pico de pressão e volume corrente. MÉTODOS: Estudo experimental com 141 profissionais de saúde (médicos, médicos residentes, fisioterapeutas, enfermeiros e técnicos de enfermagem) que ventilaram um pulmão artificial ajustado para reproduzir o pulmão de um recém-nascido a termo, utilizando ventilador pulmonar manual autoinflável com cinco diferentes modos de manuseio: com as duas mãos (10 dedos) e com uma mão, utilizando cinco, quatro, três e dois dedos. Valores de pico de pressão e volume corrente foram registrados pelo pulmão artificial. RESULTADOS: Observou-se alta variabilidade nos valores das variáveis estudadas. A média do pico de pressão foi 39,73 cmH2O (IC95 por cento 37,32-42,13), variando de 2,5 a 106,3 cmH2O, e do volume foi 39,56 mL (IC95 por cento 36,86-42,25), variando de 4 a 88 mL. Não foi detectada influência significativa da profissão sobre as variáveis estudadas (p > 0,05). Observou-se que o modo de manuseio do ventilador pulmonar manual influenciou significativamente o pico de pressão e o volume corrente (p < 0,0001), que foram maiores quando o ventilador foi manuseado com as duas mãos. CONCLUSÃO: A maioria dos profissionais gerou pico de pressão e volume corrente muito elevados, principalmente quando foram empregadas as duas mãos para ventilar, aumentando o risco de barotrauma e volutrauma. Por outro lado, parte dos profissionais forneceu pressões e volumes tão baixos que seriam insuficientes para a adequada expansão pulmonar, podendo levar a hipoventilação. Tais riscos não dependeram da formação profissional.
ABSTRACT
OBJECTIVE: To evaluate how different ways of handling the neonatal self-inflating bag influence peak pressure and tidal volume. METHODS: This is an experimental study involving 141 different professionals (physicians, resident physicians, physiotherapists, nurses, and nursing technicians), who ventilated an artificial lung, adjusted to simulate the lung of a term neonate, using a self-inflating bag. Each professional handled the ventilator in five different ways: a) using both hands (10 fingers); and, with only one hand, b) five fingers, c) four fingers, d) three fingers, and e) two fingers. Peak pressure and tidal volume data were recorded by the artificial lung equipment. RESULTS: Both variables showed high variability, from 2.5 to 106.3 cmH2O (mean = 39.73 cmH2O; 95 percentCI 37.32-42.13) for peak pressure, and from to 4 to 88 mL (mean = 39.56 mL; 95 percentCI 36.86-42.25) for tidal volume. There was no significant influence of the profession on any of the variables (p > 0.05). However, bag handling significantly influenced both peak pressure and tidal volume (p < 0.0001), which were higher when the operator used both hands. CONCLUSION: The results indicate that most professionals delivered excessively high peak pressures and tidal volumes, which could increase the risk of barotrauma and volutrauma, especially when both hands were used to ventilate. On the other hand, a small number of professionals delivered insufficient pressure and volume for adequate lung expansion and ventilation. The delivery of inadequate ventilation was not dependent on profession.
Asunto(s)

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Presión / Órganos Artificiales / Respiración Artificial / Pulmón Límite: Humanos / Recién Nacido Idioma: Inglés / Portugués Revista: J. pediatr. (Rio J.) Asunto de la revista: Pediatría Año: 2009 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: UNICAMP/BR / Universidade Estadual de Campinas/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Presión / Órganos Artificiales / Respiración Artificial / Pulmón Límite: Humanos / Recién Nacido Idioma: Inglés / Portugués Revista: J. pediatr. (Rio J.) Asunto de la revista: Pediatría Año: 2009 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: UNICAMP/BR / Universidade Estadual de Campinas/BR