Reducción del ruido en imágenes de tomografía computarizada usando un filtro bilateral anisotrópico / Noise reduction in computed tomography images using an anisotropic bilateral filter
Ramirez Giraldo, Juan Carlos; Fletcher, Joel J; McCollough, Cynthia H.
  • Ramirez Giraldo, Juan Carlos; Mayo Clinic. CT Clinical Innovation Center. Department of Physhiology and Biomedical Engineering. Rochester. US
  • Fletcher, Joel J; Mayo Clinic. CT Clinical Innovation Center. Rochester. US
  • McCollough, Cynthia H; Mayo Clinic. CT Clinical Innovation Center. Rochester. US
Rev. ing. bioméd ; 4(7): 55-62, ene.- jun. 2010. ilus, tab, graf
Artículo en Español | LILACS | ID: lil-590324
RESUMEN
En este trabajo se propone el uso de un filtro bilateral anisotrópico (FBA) para reducir el ruido en imágenes de tomografía computarizada. El FBA fue implementado en una versión tridimensional que permite ajustar los parámetros del filtro dependiendo de la resolución de las imágenes en cada uno de los ejes. La utilidad del FBA se demostró con un fantoma estándar, que se escaneó, inicialmente, utilizando una dosis de radiación referencia de 240 mAs, y, seguidamente, con dosis del 50 % y 25% de la de referencia. Asimismo, se procesaron y analizaron dos casos clínicos correspondientes a una tomografía abdominal y otra de tórax, ambas utilizando una inyección intravenosa de medio de contraste. Se encontró que el FBA permite mantener una mejor relación entre el ruido, la resolución espacial y la detectabilidad de bajos contrastes, cuando se le compara con el método tradicional de retroproyección filtrada que utilizan los escáneres de tomografía clínicos. Los resultados del fantoma, sugieren que es posible reducir las dosis de radiación hasta en un 50% sin afectar la resolución espacial o la detectabilidad de bajos contrastes, cuando se le compara con la dosis de referencia. Los estudios clínicos, revelaron que el FBA puede disminuir el ruido de las imágenes y aún garantizar una calidad adecuada para el diagnóstico. Estudios clínicos prospectivos, son necesarios para demostrar que la disminución del ruido puede permitir una reducción significativa de las dosis de radiación.
ABSTRACT
This work proposes the use of an anisotropic bilateral filter (ABF) to reduce the noise in computed tomography images. The FBA is implemented in the tridimensional space, allowing the adjustment of filter parameters depending on the image resolution in each axis. The utility of the FBA was demonstrated using a standard image quality phantom which was scanned using a 240 mAs reference dose, and subsequently with 50 % and 25 % of the reference dose. Additionally, two clinical cases were processed, corresponding to routine clinical, intravenous contrast-enhanced abdomen and thorax scans. The phantom study found that the FBA keeps a better tradeoff between noise, spatial resolution, and low contrast detectability, when it is compared to traditional filtered backprojection reconstructions, routinely employed by current computed tomography scanners. Hence, results in the phantom suggest it is possible to reduce radiation dose by at least 50% without affecting spatial resolution or low contrast detectability. The clinical studies revealed that the FBA can decrease image noise and still provide enough information for adequate diagnosis. Prospective clinical studies are necessary to demonstrate whether or not the observed noise reduction would allow a significant decrease in radiation dose.
Asunto(s)

Buscar en Google
Índice: LILACS (Américas) Asunto principal: Radiación Ionizante / Tomografía Computarizada por Rayos X / Filtros / Control de la Radiación / Monitoreo del Ruido Límite: Humanos Idioma: Español Revista: Rev. ing. bioméd Asunto de la revista: Biotecnologia / Ingenieria Biomédica Año: 2010 Tipo del documento: Artículo País de afiliación: Estados Unidos Institución/País de afiliación: Mayo Clinic/US

Similares

MEDLINE

...
LILACS

LIS

Buscar en Google
Índice: LILACS (Américas) Asunto principal: Radiación Ionizante / Tomografía Computarizada por Rayos X / Filtros / Control de la Radiación / Monitoreo del Ruido Límite: Humanos Idioma: Español Revista: Rev. ing. bioméd Asunto de la revista: Biotecnologia / Ingenieria Biomédica Año: 2010 Tipo del documento: Artículo País de afiliación: Estados Unidos Institución/País de afiliación: Mayo Clinic/US