Effect of pre-flaring and file size on the accuracy of two electronic apex locators
Brito-Júnior, Manoel; Camilo, Carla Cristina; Moreira-Júnior, Gil; Pecora, Jesus Djalma; Sousa-Neto, Manoel Damião.
  • Brito-Júnior, Manoel; University of São Paulo. School of Dentistry. Department of Restorative Dentistry. Ribeirão Preto. BR
  • Camilo, Carla Cristina; University of São Paulo. School of Dentistry. Department of Restorative Dentistry. Ribeirão Preto. BR
  • Moreira-Júnior, Gil; University of Itaúna. Dental School. Itaúna. BR
  • Pecora, Jesus Djalma; University of São Paulo. School of Dentistry. Ribeirão Preto. BR
  • Sousa-Neto, Manoel Damião; University of São Paulo. School of Dentistry. Ribeirão Preto. BR
J. appl. oral sci ; 20(5): 538-543, Sept.-Oct. 2012. ilus, graf
Artículo en Inglés | LILACS | ID: lil-654918
ABSTRACT

OBJECTIVE:

This ex vivo study evaluated the effect of pre-flaring and file size on the accuracy of the Root ZX and Novapex electronic apex locators (EALs). MATERIAL AND

METHODS:

The actual working length (WL) was set 1 mm short of the apical foramen in the palatal root canals of 24 extracted maxillary molars. The teeth were embedded in an alginate mold, and two examiners performed the electronic measurements using #10, #15, and #20 K-files. The files were inserted into the root canals until the "0.0" or "APEX" signals were observed on the LED or display screens for the Novapex and Root ZX, respectively, retracting to the 1.0 mark. The measurements were repeated after the preflaring using the S1 and SX Pro-Taper instruments. Two measurements were performed for each condition and the means were used. Intra-class correlation coefficients (ICCs) were calculated to verify the intra- and inter-examiner agreement. The mean differences between the WL and electronic length values were analyzed by the three-way ANOVA test (p<0.05).

RESULTS:

ICCs were high (>0.8) and the results demonstrated a similar accuracy for both EALs (p>0.05). Statistically significant accurate measurements were verified in the pre-flared canals, except for the Novapex using a #20 K-file.

CONCLUSIONS:

The tested EALs showed acceptable accuracy, whereas the pre-flaring procedure revealed a more significant effect than the used file size.
Asunto(s)


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Preparación del Conducto Radicular / Ápice del Diente / Instrumentos Dentales / Cavidad Pulpar / Odontometría Tipo de estudio: Estudios de evaluación Límite: Humanos Idioma: Inglés Revista: J. appl. oral sci Asunto de la revista: Odontología Año: 2012 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: University of Itaúna/BR / University of São Paulo/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Preparación del Conducto Radicular / Ápice del Diente / Instrumentos Dentales / Cavidad Pulpar / Odontometría Tipo de estudio: Estudios de evaluación Límite: Humanos Idioma: Inglés Revista: J. appl. oral sci Asunto de la revista: Odontología Año: 2012 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: University of Itaúna/BR / University of São Paulo/BR