Estudo experimental comparativo entre a derivação cavo-pulmonar associada ao suporte circulatório esquerdo e o suporte circulatório biventricular na falência aguda biventricular / Comparative experimental study between cavopulmonary anastomosis associated with left ventricular assist device support and biventricular circulatory assistance in acute biventricular failure
São Paulo; s.n; 2012. [143] p. tab, graf, ilus.
Thesis
en Pt
| LILACS
| ID: lil-664746
Biblioteca responsable:
BR66.1
Ubicación: BR66.1; W4.DB8/SP.USP/FM-2, S236es, 2012
RESUMO
A falência do ventrículo direito (VD), durante o emprego de dispositivos de assistência circulatória do ventrículo esquerdo (DAVE), pode resultar em severo compromisso hemodinâmico com altas taxas de mortalidade. Objetivo:
Este estudo avaliou o desempenho hemodinâmico e as alterações miocárdicas decorrentes do emprego de dispositivo de assistência ventricular esquerda, associado ou não à descompressão do ventrículo direito (VD) através de derivação cavo-pulmonar.Estes achados foram comparados com o emprego de assistência circulatória esquerda isolada e com a assistência circulatória biventricular, em modelo experimental de falência cardíaca.Métodos:
O suporte ventricular esquerdo foi instalado por meio de bomba centrífuga em vinte e um suínos (25-32 kg) com falência biventricular severa obtida pela indução de ritmo de fibrilação ventricular (FV). Os animais foram randomizados em três grupos. No grupo Controle, foi implantado apenas o DAVE. No grupo Derivação, além do DAVE foi realizada a cirurgia de Glenn modificada, com anastomose de tubo de Politetrafluoroetileno expandido (PTFE) n° 16, entre a veia cava superior e o tronco da artéria pulmonar. No grupo Biventricular, foram instaladas duas bombas centrífugas em paralelo, uma do lado esquerdo e outra do lado direito do coração. Foram monitoradas as pressões intracavitárias no momento prévio à indução de FV e a cada 30 minutos de assistência até o limite de 180 minutos. Lactato venoso e dosagens de citocinas inflamatórias também foram coletados em momentos distintos no inicio da FV e a Intervalos regulares de uma hora até o final do experimento Amostras do endocárdio dos dois ventrículos foram coletadas e analisadas por meio de microscopia óptica e eletrônicaResultados:
manutenção do ritmo de FV foi responsável por detenoração hernodinâmica Importante no grupo Controle e no grupo Derivação A diferença observada entre o fluxo do...ABSTRACT
Background Right ventricular (RV) failure durinq left ventricular assist device (LVAD) support can result in severe hernodynarnic cornpromise with hiqh mortality. Objective:
This study investiqated the acute effects of cavopulmonary anastomosis on L VAD performance and riqht ventricular myocardial compromise in comparison with isolated LVAD support and biventricular circulatory support in a model of severe cardiac failure.Methods:
LVAD support was performed by means of centrituqa pump implantation in 21 anesthetized guinea piqs (25-32 kg) with severe biventricular failure achieved by ventricular fibrillation (VF) induction. Animais were randomized in three groups In the Control group, isolated LVAD support was performed In the Cavopulmonary group, cavopulmonary anastomosis was established by connecting the superior vena cava ano pulmonary main artery with a short segment of expanded politrafluorotylen no. 16 (rnodified Glenn shunt). In the Biventricular group, biventricular assist device support was established by means of two parallel centrifugal pumps, one of them on the riqht-hand side of the heart and the other on the left-hand side. They were maintened under circulatory support and hemodynamic monitoring for 3h. Blood sample was collected before VF and in intervals of 30 minutes until the end of the surgery. Venous lactate and cytokines serum levels derterminations were also obtained, before VF and each hour. Endocardium samples of the septum and two ventricles were collected and analyzed by optics and electronic in microscopy.Results:
Ventricular fibrillation maintenance was responsible for acute LVAD performance impairment after mino in the Control and Cavopulmonary groups. Cavopulmonary anastomosis resulted in non-significant improvement of LVAD pump flow in relation to control group (+55±14 ml/kg/min, p=0.072), while animais under biventricular support maintained higher LVAD flow performance (+93±17 ml/kg/min, p=0.012)...
Texto completo:
1
Índice:
LILACS
Asunto principal:
Porcinos
/
Corazón Auxiliar
/
Disfunción Ventricular Derecha
/
Puente Cardíaco Derecho
/
Insuficiencia Cardíaca
Tipo de estudio:
Prognostic_studies
/
Risk_factors_studies
Límite:
Animals
Idioma:
Pt
Año:
2012
Tipo del documento:
Thesis