Precisão e confiabilidade de um novo localizador foraminal eletrônico: estudo in vivo / Accuracy and reliability of a new electronic apex locator: an in-vivo study
Chita, Jair Jatobá; Silva, Pedro Gregol da; Pereira, Key Fabiano Souza; Onoda, Helio Katsuya; Borba Junior, José de Camargo; Ramos, Carlos Alberto Spironelli.
  • Chita, Jair Jatobá; Universidade Federal de Mato Grosso do Sul. Faculdade de Odontologia Prof. Albino Coimbra Filho. Cuiabá. BR
  • Silva, Pedro Gregol da; Universidade Federal de Mato Grosso do Sul. Faculdade de Odontologia Prof. Albino Coimbra Filho. Cuiabá. BR
  • Pereira, Key Fabiano Souza; Universidade Federal de Mato Grosso do Sul. Faculdade de Odontologia Prof. Albino Coimbra Filho. Cuiabá. BR
  • Onoda, Helio Katsuya; Universidade Federal de Mato Grosso do Sul. Faculdade de Odontologia Prof. Albino Coimbra Filho. Cuiabá. BR
  • Borba Junior, José de Camargo; Universidade para Desenvolvimento do Pantanal. Cuiabá. BR
  • Ramos, Carlos Alberto Spironelli; Universidade Estadual de Londrina. Londrina. BR
Artículo en Portugués | LILACS, BBO | ID: lil-673926
RESUMO

Objetivo:

Verificar in vivo a precisão e confiabilidade de leitura do localizador foraminal eletrônico Joypex 5® utilizando o microscópio eletrônico de varredura na aferição das medidas obtidas pelo aparelho.

Método:

Foram triados pacientes com indicação para extração por motivos ortodônticos e periodontais, o que resultou em amostra de 14 canais. Os dentes foram previamente radiografados com a intenção de detectar perfurações, tratamentos endodônticos prévios e calcificações. Realizadas as aberturas coronárias, os terços cervical e médio foram preparados e procederam-se as leituras no ponto correspondente no display do aparelho ao forame apical. Após a obtenção da medida a lima foi removida e subtraiu-se 1mm do comprimento lido em paquímetro digital para, então, fixar o instrumento e realizar a exodontia. Na seqüência foi realizado um desgaste em uma das paredes da região apical, objetivando visualizar a ponta do instrumento e a continuidade do canal até a real saída do forame apical. A distância entre a ponta da lima e a real saída foraminal foi medida com o auxílio de microscopia eletrônica de varredura.

Resultados:

A média das medidas foi 0,87 mm (DP 0,42mm). O teste T para amostras independentes revelou que os dados mostraram-se semelhantes (p > 0,05) entre os valores experimentais encontrados (da ponta do instrumento ao forame apical) e o valor hipotético testado de 1 mm.

Conclusões:

O estudo concluiu que o localizador foraminal eletrônico Joypex 5® demonstrou ser preciso e confiável na determinação de um comprimento de trabalho eficiente e seguro para o tratamento endodôntico. de trabalho eficiente e seguro para o tratamento endodôntico.
ABSTRACT

Objective:

To evaluate in vivo the measurement accuracy and reliability of the electronic apex locator Joypex 5® using a scanning electron microscope to assess the measurements obtained with the tested device.

Method:

Patients with indication for tooth extraction for orthodontic and periodontal reasons were screened, providing a sample of 14 root canals. The teeth were first radiographed for detecting perforations, previous endodontic treatments and calcifications. After preparation of access cavities, the coronal and middle thirds were prepared and the electronic measurements were made on the point corresponding to the apical foramen of the apex locator's display. After the electronic reading, the file was removed from the canal, 1 mm was subtracted from the measurement obtained using a digital caliper, and the apex locator was fixed for extraction of the tooth. Next, grinding was performed on an apical wall in order to visualize the tip of the file and canal integrity up to the actual apical foramen opening. The distance between the file tip and the actual apical foramen opening was measured by scanning electron microscopy.

Results:

The mean of the obtained measurements was 0.87 mm (DP 0.42 mm). The t test for independent samples revealed statistical similarity (p>0.05) among the obtained values (from the file tip to actual apical foramen) and the hypothetical 1 mm value.

Conclusion:

The Joypex 5® electronic apex locator was found to be accurate and reliable for determination of an efficient and safe working length for endodontic treatment.
Asunto(s)

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Microscopía Electrónica de Rastreo / Cavidad Pulpar / Endodoncia / Odontometría Límite: Humanos País/Región como asunto: America del Sur / Brasil Idioma: Portugués Revista: Pesqui. bras. odontopediatria clín. integr Asunto de la revista: Odontología Año: 2012 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Universidade Estadual de Londrina/BR / Universidade Federal de Mato Grosso do Sul/BR / Universidade para Desenvolvimento do Pantanal/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Microscopía Electrónica de Rastreo / Cavidad Pulpar / Endodoncia / Odontometría Límite: Humanos País/Región como asunto: America del Sur / Brasil Idioma: Portugués Revista: Pesqui. bras. odontopediatria clín. integr Asunto de la revista: Odontología Año: 2012 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Universidade Estadual de Londrina/BR / Universidade Federal de Mato Grosso do Sul/BR / Universidade para Desenvolvimento do Pantanal/BR