Effectiveness of epicardial atrial pacing using a bipolar steroid-eluting endocardial lead with active fixation in an experimental model
Bueno, Sávia Christina Pereira; Tamaki, Wagner Tetsuji; Silva, Marcelo Fiorelli Alexandrino da; Zambolim, Cristiane Maciel; Silva, Kátia Regina da; Martinelli Filho, Martino; Costa, Roberto.
  • Bueno, Sávia Christina Pereira; USP. Clinics Hospital. Sao Paulo. BR
  • Tamaki, Wagner Tetsuji; USP. Clinics Hospital. Sao Paulo. BR
  • Silva, Marcelo Fiorelli Alexandrino da; USP. Clinics Hospital. Sao Paulo. BR
  • Zambolim, Cristiane Maciel; USP. Clinics Hospital. Sao Paulo. BR
  • Silva, Kátia Regina da; USP. Clinics Hospital. Sao Paulo. BR
  • Martinelli Filho, Martino; USP. Clinics Hospital. Sao Paulo. BR
  • Costa, Roberto; USP. Clinics Hospital. Sao Paulo. BR
Acta cir. bras ; 28(7): 543-550, July 2013. ilus, graf, tab
Artículo en Inglés | LILACS | ID: lil-679088
ABSTRACT

PURPOSE:

To assess the effectiveness of bipolar epicardial atrial pacing using an active fixation bipolar endocardial lead implanted on the atrial surface in an experimental model.

METHODS:

A total of ten Large White adult pigs underwent pacemaker implantation under general anesthesia. Atrial pacing and sensing parameters were obtained at the procedure, immediate postoperative period and on the 7th and the 30th postoperative in unipolar and bipolar configurations.

RESULTS:

All procedures were successfully performed. There were no perioperative complications and no early deaths. Atrial pacing and sensing parameters for both unipolar and bipolar modes remained stable throughout the study. We observed a progressive increase in atrial thresholds, ranging from 0.49 ± 0.35 (at implantation) to 1.86 ± 1.31 volts (30th postoperative day), in unipolar mode. Atrial impedance measurements decreased slightly over time, ranging from 486.80 ± 126.35 Ohms (at implantation) to 385.0 ± 80.52 Ohms (30th postoperative day). Atrial sensing measures remained stable from the immediate postoperative period until the end of the study.

CONCLUSION:

The bipolar active fixation endocardial lead implanted epicardially can provide stable conditions of pacing and sensing parameters throughout the postoperative follow-up.
Asunto(s)


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Marcapaso Artificial / Esteroides / Estimulación Cardíaca Artificial / Electrodos Implantados / Diseño de Equipo / Atrios Cardíacos Tipo de estudio: Estudios de evaluación / Estudio pronóstico Límite: Animales Idioma: Inglés Revista: Acta cir. bras Asunto de la revista: Cirugía General / Procedimentos Cir£rgicos Operat¢rios Año: 2013 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: USP/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Marcapaso Artificial / Esteroides / Estimulación Cardíaca Artificial / Electrodos Implantados / Diseño de Equipo / Atrios Cardíacos Tipo de estudio: Estudios de evaluación / Estudio pronóstico Límite: Animales Idioma: Inglés Revista: Acta cir. bras Asunto de la revista: Cirugía General / Procedimentos Cir£rgicos Operat¢rios Año: 2013 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: USP/BR