Avaliação ultra-estrutural, dimensional e mecânica de dois sistemas de fixação interna rígida de 2 mm / Ultra-structural dimensional and mechanical evaluation of two 2mm systems of internal rigid fixation
Rocha Júnior, Hernando Valentim da; Padovan, Luis Eduardo Marques; Oliveira, Rogerio Belle de; Assis, Diogo Souza Ferreira Rubim de; Duarte, Marco Antonio Hungaro.
Afiliación
  • Rocha Júnior, Hernando Valentim da; Universidade do Sagrado Coração. Bauru. BR
  • Padovan, Luis Eduardo Marques; Universidade do Sagrado Coração. Bauru. BR
  • Oliveira, Rogerio Belle de; Pontifícia Universidade Católica do Rio Grande do Sul. Porto Alegre. BR
  • Assis, Diogo Souza Ferreira Rubim de; Instituto Florence Ensino Superior. São Luis. BR
  • Duarte, Marco Antonio Hungaro; Universidade de São Paulo. Bauru. BR
Rev. Salusvita (Online) ; 32(2)2013. tab, graf, ilus
Article en Pt | LILACS | ID: lil-698380
Biblioteca responsable: BR275.1
RESUMO
Introdução e

objetivo:

este trabalho teve como objetivo comparar dois materiais de fixação interna rígida utilizados em cirurgia maxilofacial, sob o ponto de vista ultra-estrutural, dimensional e mecânico.

Materiais e Métodos:

foram comparadas amostras de miniplacas de 2.0mm Neoortho® e Synthes®. As amostras foram avaliadas por Microscopia Eletrônica de Varredura (MEV) e Espectroscopia por Dispersão de Energia (EDS) para avaliação superficial e identificação de contaminantes. Avaliação macroscópica foi feita com paquímetros digitais e com micrômetro. Miniplacas foram submetidas a ensaio de acordo com as normas da American Society for Testing and Materials, Norma F382-99.

Resultado:

Como resultado do exame pelo MEV e EDS, observa-se a presença de contaminantes, manchamentos e rebarbas em ambas as amostras. Na análise dimensional, observou-se variabilidade nas medidas entre as marcas e maior variabilidade de dimensão nas miniplacas Neoortho®. Ao ensaio de flexão, observou-se discrepância entre as amostras, sendo a Synthes® com maior homogeneidade, mas ambas dentro do intervalo preconizado pela ASTM.

Conclusão:

a MEV demonstrou superfícies de titânio homogêneas e com debris de superfície oriundos da fabricação e manipulação dos materiais. Dimensionalmente as miniplacas apresentam variações quando comparadas, e estas podem influenciar diretamente os resultados do dos ensaios mecânicos, sendo que mesmo com as variações ambas estão de acordo com a norma da ASTM.
ABSTRACT
Introduction and

objective:

this study aimed to compare two rigid internal fixation materials used in maxillofacial surgery, from the ultra-structural, dimensional and mechanical points of view. Materials and

Methods:

we have compared samples of 2.0 mm miniplates produced by Neoortho® and Synthes®. The samples were analyzed by Scanning Electron Microscopy (SEM) and Energy Dispersive Spectroscopy (EDS) for evaluation and identification of surface contaminants. Macroscopic evaluation was made with digital calipers and micrometer. The miniplates were subjected totesting in accordance with the F382-99 standards of the American Society for Testing and Materials Standard.

Result:

examination by SEM and EDS revealed the presence of contaminants, staining and burrs in both samples. In dimensional analysis, there was variability in measurements between brands and greater variability in size of Neoortho® miniplates. In the bending test, it was observed discrepancy between the samples, showing the Synthes® greater uniformity, but both within the range recommended by ASTM.

Conclusion:

the SEM showed homogeneous titanium surfaces and surface debris deriving from the manufacture and handling of materials. From the dimensional point of view, the miniplates showed variations when compared to each other, and this may directly influence the results of the mechanical tests. However, even with variations both are in agreement with the standard ASTM..