Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory
Bitencourt, Eberson Damiao dos Santos de; Voegeli, Carlos Franco; Onzi, Gabriela dos Santos; Boscato, Sara Cardoso; Ghem, Carine; Munhoz, Terezinha.
  • Bitencourt, Eberson Damiao dos Santos de; Pontificia Universidade Catolica do Rio Grande do Sul. Porto Alegre. BR
  • Voegeli, Carlos Franco; Pontificia Universidade Catolica do Rio Grande do Sul. Porto Alegre. BR
  • Onzi, Gabriela dos Santos; Pontificia Universidade Catolica do Rio Grande do Sul. Porto Alegre. BR
  • Boscato, Sara Cardoso; Pontificia Universidade Catolica do Rio Grande do Sul. Porto Alegre. BR
  • Ghem, Carine; Pontificia Universidade Catolica do Rio Grande do Sul. Porto Alegre. BR
  • Munhoz, Terezinha; Pontificia Universidade Catolica do Rio Grande do Sul. Porto Alegre. BR
Rev. bras. hematol. hemoter ; 35(6): 404-408, 2013. tab, graf
Artículo en Inglés | LILACS | ID: lil-699995
ABSTRACT

Background:

The speed and quality of information have become essential items in the release of laboratory reports. The Sysmex®SP1000-I device has been developed to prepare and stain smear slides. However, for a device to be cleared for use in the laboratory routine it must pass through a validation process.

Objective:

To evaluate the performance and reliability of the Sysmex® SP-1000i slide preparer-stainer incorporated into the routine of a hospital laboratory in Porto Alegre.

Methods:

Peripheral blood samples of patients attending the laboratory for ambulatory exams with leukocyte counts between 7000/°L and 12,000/°L were evaluated, independent of gender and age. Two slides were prepared for each sample using the Sysmex® SP-1000i equipment; one of the slides was used to perform quality control tests using the CellaVision® DM96 device, and the other slide was used to compare pre-classification by the same device and the classification performed by a pharmacist-biochemist.

Results:

The results of all the slides used as controls were acceptable according to the quality control test as established by the manufacturer of the device. In the comparison between the automated pre-classification and the classification made by the professional, there was an acceptable variation in the differential counts of leukocytes for 90% of the analyzed slides. Pearson correlation coefficient showed a strong correlation for band neutrophils (r = 0.802; p-value < 0.001), segmented neutrophils (r = 0.963; p-value < 0.001), eosinophils (r = 0.958; p-value < 0.001), lymphocytes (r = 0.985; p-value < 0.001) and atypical lymphocytes (r = 0.866; p-value < 0.001) using both methods. The red blood cell analysis was adequate for all slides analyzed by the equipment and by the professional.

Conclusion:

The new Sysmex®SP1000-i methodology was found to be reliable, fast and safe for the routines of medium ...
Asunto(s)


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Recuento de Células Sanguíneas / Equipo para Diagnóstico / Estudio de Validación / Automatización de Laboratorios / Citometría de Flujo Idioma: Inglés Revista: Rev. bras. hematol. hemoter Asunto de la revista: Hematología Año: 2013 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Pontificia Universidade Catolica do Rio Grande do Sul/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Recuento de Células Sanguíneas / Equipo para Diagnóstico / Estudio de Validación / Automatización de Laboratorios / Citometría de Flujo Idioma: Inglés Revista: Rev. bras. hematol. hemoter Asunto de la revista: Hematología Año: 2013 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Pontificia Universidade Catolica do Rio Grande do Sul/BR