Caracterização micrométrica das superfícies dos implantes das cinco maiores companhias do mercado brasileiro. Parte I: implantes da Neodent / Characterization of implants surface of the five largest companies in the Brazilian market, on micrometric leveI. Part I: Neodent implants
Rosa, Marcio Borges; Albrektsson, Tomas; Francischone, Carlos Eduardo; Schwartz Filho, Humberto Osvaldo; Wennerberg, Ann.
Afiliación
  • Rosa, Marcio Borges; São Leopoldo Mandic. Campinas. BR
  • Albrektsson, Tomas; Universidade de Gotemburgo. Departamento de Biomateriais. SE
  • Francischone, Carlos Eduardo; USP. Faculdade de Odontologia de Bauru. Bauru. BR
  • Schwartz Filho, Humberto Osvaldo; Unesp. Araraquara. BR
  • Wennerberg, Ann; Faculdade de Malmo. SE
Dent. press implantol ; 6(1): 76-87, jan.-mar. 2012. ilus, tab, graf
Article en Pt | LILACS, BBO | ID: lil-700330
Biblioteca responsable: BR1366.1
RESUMO

Introdução:

a qualidade da interface entre osso-implante é influenciada diretamente pela rugosidade da superfície do implante e uma rugosidade média, com Sa entre 1 e 2µm, tem demonstrado melhores resultados clínicos e laboratoriais. No Brasil, são instalados mais de dois milhões de implantes por ano, onde 79% são fabricados por empresas nacionais. Porém, muito pouco é divulgado ou se conhece sobre a caracterização das superfícies desses implantes, a nível micrométrico. Este estudo visa avaliar e caracterizar, numericamente, a superfície dos implantes da empresa Neodent, uma das cinco maiores empresas do mercado brasileiro.

Métodos:

foram avaliados três implantes, comprados diretamente no mercado, de três desenhos da companhia e de diferentes lotes, através de um interferômetro de luz. Foram realizados nove medidas, escolhidas aleatoriamente, para cada unidade, sendo três nos topos, três nos vales e três nos flancos das roscas. O mesmo padrão foi seguido para avaliação através microscópio eletrônico de varredura.

Resultados:

de uma forma geral, os implantes analisados desta companhia, apresentaram valores de Sa de 0,75µm, 0,67µm e 0,65µm, respectivamente para cada desenho analisado. Na comparação entre os lotes, todos os desenhos apresentaram diferenças estatisticamente significativa, entre pelo menos um lote em relação aos outros.

Conclusões:

Os valores de rugosidade encontrados, classificam as superfícies, dos três implantes avaliados, como minimamente rugosas.
ABSTRACT

Introduction:

The quality of the bone-implant interface is directly influenced by implant surface roughness and a roughness average, with the Sa between 1 to 2 µm, has demonstrated better clinical and laboratary results. In Brazil, are installed more than two million implants per year, where 79% are manufactured by domestic companies. However, very little is known or published about the characterization of surfaces of these implants, on the micrometer level. The aims of this study are to evaluate and characterize numerically the surface of the implants of Neodent company, one of the five largest companies in the Brazilian market. Materials and

methods:

were evaluated a total of 9 implants, purchased directly on the market, of 3 different designs and different batches of the company, using a light interferometer. Were performed 9 measurements randomly chosen for each unit, 3 on the tops, 3 on the valleys and 3 on the flanks of the threads. The same pattern was followed for evaluation by scanning electron microscope.

Results:

in general, implants analyzed in this company, showed Sa values of 0.75 µm, 0.67 µm and 0.65µm, respectively, for each design. Comparing the batches, all designs presented statistically significant differences between at least one batches in relation to other.

Conclusions:

the roughness values found, classify the surfaces of the three implants evaluated as minimally rough.
Asunto(s)