Modificações na angulação coronária após implante de suporte vascular bioabsorvível e de stents de cromo-cobalto e aço inoxidável / Changes in coronary angulation after bioresorbable vascular scaffold and cobalt-chromium and stainless steel stent implantation
Silva, Mateus Veloso e; Costa Junior, J. Ribamar; Abizaid, Alexandre; Staico, Rodolfo; Taiguara, Danillo; Borghi Junior, Tarcísio Campostrini; Costa, Ricardo; Chamié, Daniel; Sousa, Amanda G. M. R; Sousa, J. Eduardo.
  • Silva, Mateus Veloso e; s.af
  • Costa Junior, J. Ribamar; Instituto Dante Pazzanese de Cardiologia. Serviço de Cardiologia Invasiva. São Paulo. BR
  • Abizaid, Alexandre; Instituto Dante Pazzanese de Cardiologia. Serviço de Cardiologia Invasiva. São Paulo. BR
  • Staico, Rodolfo; Instituto Dante Pazzanese de Cardiologia. Serviço de Cardiologia Invasiva. São Paulo. BR
  • Taiguara, Danillo; s.af
  • Borghi Junior, Tarcísio Campostrini; s.af
  • Costa, Ricardo; Instituto Dante Pazzanese de Cardiologia. Serviço de Cardiologia Invasiva. São Paulo. BR
  • Chamié, Daniel; Instituto Dante Pazzanese de Cardiologia. Serviço de Cardiologia Invasiva. São Paulo. BR
  • Sousa, Amanda G. M. R; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Sousa, J. Eduardo; Instituto Dante Pazzanese de Cardiologia. Centro de Intervenções em Doenças Estruturais do Coração. São Paulo. BR
Rev. bras. cardiol. invasiva ; 21(4): 332-337, out.-dez. 2013. ilus, tab
Artículo en Portugués | LILACS | ID: lil-703684
RESUMO

INTRODUÇÃO:

A conformabilidade do stent, definida como a adaptação da prótese à forma natural do vaso, é a principal responsável pelas alterações geométricas produzidas após o implante do dispositivo, sendo influenciada pelo material e pelo desenho do stent. Ela pode ser aferida medindo-se as modificações da curvatura e a angulação do segmento tratado após o implante do stent. O objetivo deste estudo foi comparar as mudanças na angulação coronária após implante do suporte vascular bioabsorvível (SVB) e das plataformas metálicas de cromo-cobalto e aço inoxidável, utilizadas em stents farmacológicos de segunda geração.

MÉTODOS:

Foram incluídos, nesta análise retrospectiva, 50 pacientes com lesões únicas, de novo, em coronárias nativas, com diâmetro entre 2,5 e 3,5 mm e extensão de até 23 mm. Vinte e cinco pacientes foram tratados com SVB e 25 pacientes com plataformas metálicas de cromo-cobalto (n = 12) ou de aço inoxidável (n = 13). A angulação foi medida usando um software de angiografia coronária quantitativa dedicado.

RESULTADOS:

A angulação do vaso modificou-se significativamente após o implante dos dispositivos. No grupo submetido ao implante de plataformas metálicas, houve maior modificação do ângulo coronário quando comparado ao tratado com SVB (41,6% vs. 14,9%; P < 0,01). Quando observado o comportamento do SVB e das diferentes plataformas metálicas, a alteração do ângulo coronário foi maior para as plataformas de aço inoxidável, seguida das plataformas de cromo-cobalto e os SVB (53,7% vs. 28,5% vs. 14,9%; P < 0,01).

CONCLUSÕES:

Nesta avaliação preliminar, o SVB produziu uma menor mudança da angulação coronária. O impacto clínico deste achado necessita ser investigado prospectivamente em uma coorte maior e mais complexa.
ABSTRACT

BACKGROUND:

The conformability of the stent, defined as the adaptation of the prosthesis to the natural shape of the vessel, is the major cause of geometrical changes after stenting and is influenced by the stent material and design. It may be assessed by measuring changes in the curvature and the angulation of the treated segment after stent implantation. The objective of this study was to compare changes in coronary angulation after implantation of the bioresorbable vascular scaffold (BVS) and cobalt-chromium and stainless steel metal platforms used in second-generation drug-eluting stents.

METHODS:

In this retrospective analysis, 50 patients with single de novo lesions in native coronary arteries and diameter between 2.5 and 3.5 mm and length up to 23 mm were included. Twenty-five patients were treated with BVS and 25 patients were treated with cobalt-chromium (n = 12) or stainless steel (n = 13) platforms. Angulation was measured using a dedicated quantitative angiography analysis software.

RESULTS:

Vessel angulation significantly changed after device implantation. In the group submitted to the implantation of metal platforms there was greater coronary angulation change when compared to the group treated with BVS (41.6% vs. 14.9%; P < 0.01). When we analyzed the performance of the BVS and the different metal platforms, coronary angulation change was greater for the stainless steel platforms, followed by cobalt-chromium platforms and the BVS (53.7% vs. 28.5% vs. 14.9%; P < 0.01).

CONCLUSIONS:

In this preliminary assessment, the BVS produced a smaller coronary angulation change. The clinical impact of this finding must be prospectively investigated in a larger and more complex cohort.
Asunto(s)


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Vasos Coronarios / Implantes Absorbibles / Stents Liberadores de Fármacos Tipo de estudio: Estudio observacional Límite: Femenino / Humanos / Masculino Idioma: Portugués Revista: Rev. bras. cardiol. invasiva Asunto de la revista: Cardiología / Cirugía General Año: 2013 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Instituto Dante Pazzanese de Cardiologia/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Vasos Coronarios / Implantes Absorbibles / Stents Liberadores de Fármacos Tipo de estudio: Estudio observacional Límite: Femenino / Humanos / Masculino Idioma: Portugués Revista: Rev. bras. cardiol. invasiva Asunto de la revista: Cardiología / Cirugía General Año: 2013 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Instituto Dante Pazzanese de Cardiologia/BR