Implante de marcapasso endocárdico em paciente com válvulas tricúspide e mitral metálicas / Implantation of endocardial pacemaker in a patient with metal tricuspid and mitral valves
RELAMPA, Rev. Lat.-Am. Marcapasso Arritm
; 27(1): 58-61, jan.-mar.2014.
Article
en Pt
| LILACS, SES-SP
| ID: lil-767322
Biblioteca responsable:
BR44.1
RESUMO
O implante de marcapasso endocárdico transvenoso é contraindicado em pacientes com válvulatricúspide mecânica. Relata-se o caso de uma mulher de 79 anos de idade, com válvulas mitral e tricúspidemetálicas para estimulação permanente devido à fibrilação atrial crônica de baixa resposta ventricular, que passoua apresentar sintomas de pré-síncope. Um cabo-eletrodo endocárdico foi colocado no ventrículo esquerdo porvia transvenosa através do seio coronariano na veia cardíaca posterolateral. O implante foi realizado através doseio coronário. Oferece um ritmo seguro e eficaz em pacientes com válvula tricúspide mecânica, eliminandoa necessidade de toracotomia para o implante epicárdico de cabo-eletrodo. Normalmente, a colocação é feitamediante toracotomia anterolateral ou esternotomia. Nesse caso, tanto a vida do eletrodo é reduzida quanto o seulimiar de comando geralmente aumenta com o tempo. O tecido em torno dos ventrículos é friável, especialmenteapós cirurgia cardíaca, e um eventual dano para os ventrículos durante a incisão pode ser fatal...
ABSTRACT
Transvenous endocardial pacemaker implantation is contraindicated in patients with a mechanicaltricuspid valve. An endocardial lead was placed in the left ventricle by transvenous approach through thecoronary sinus in the posterolateral cardiac vein in 79 year-old woman with metal mitral and tricuspid valve forpermanent pacing due to chronic atrial fibrillation with low ventricular response and nearsyncope symptoms.It was implanted through the coronary sinus providing a safe and effective rhythm in patients with mechanicaltricuspid valve, thereby eliminating the need for thoracotomy to implant an epicardial electrode. Transvenousendocardial pacemaker implantation is not indicated in patients with a mechanical tricuspid valve. Typically, theyare implanted by anterolateral thoracotomy or sternotomy. However, in this case, the lead mean life is reduced,and its threshold will usually increase over time. The tissue around the ventricles is friable, especially after cardiacsurgery, and an eventual ventricular damage during incision may be fatal...
Palabras clave
Texto completo:
1
Índice:
LILACS
Asunto principal:
Marcapaso Artificial
/
Estenosis de la Válvula Tricúspide
/
Estenosis de la Válvula Mitral
Tipo de estudio:
Diagnostic_studies
Límite:
Aged
/
Female
/
Humans
Idioma:
Pt
Revista:
RELAMPA, Rev. Lat.-Am. Marcapasso Arritm
Asunto de la revista:
CARDIOLOGIA
Año:
2014
Tipo del documento:
Article