Considerações sobre o modelo experimental de hipertensão intracranianae avaliação do sistema de microchip para monitoração de pressão epidural / Experimental model of intracranial hypertension and evaluation of the microchip system for epidural pressure monitoring of epidural pressure
Arq. bras. neurocir
;
33(4): 300-305, dez. 2014. ilus, tab
Artículo
en Portugués
| LILACS
| ID: lil-782246
RESUMO
Objetivo:
Neste trabalho temos o objetivo de avaliar a acura?cia do sistema de aferic?a?o da pressão intracraniana (PIC) epidural com microchip. Me?todos Foram estudados 27 sui?nos sob anestesia geraI, devidamente assistidos com monitoração ventilatória e hemodina?mica. Durante o experimento foi reproduzido um processo expansivo intracerebral programado no lobo frontal direito. O experimento constou de tre?s grupos (A, B e C) com hipertensão intracraniana gerada com balão reproduzindoum hematoma intracerebral. Em todos os grupos foram calibrados os para?metros normais os dois sistemas de PIC foram comparados e estudados quanto a? correlação dos valores aferidos.Resultados:
O comportamento médio da PIC ao longo dos momentos de avaliac?a?o foi estatisticamente diferente entre os grupos (p < 0,001). A reprodução de ressangramento resultou em elevac?a?o significativa da PIC (p < 0,001). Avaliando-se a acura?cia comparativa geral, verificou-se um coeficiente de correlação intraclasse de 0,8. Conclusa?o O modelo de hipertensa?o intracraniana por bala?o em sui?nos e? facti?vel e confia?vel na gerac?a?o de hipertensa?o intracraniana. O sistema de aferic?a?o de pressa?o intracranianaepidural apresenta elevado coeficiente de correlac?a?o com o sistema de aferic?a?o parenquimatoso na avaliac?a?o global.ABSTRACT
Objective:
In this paper we aim to evaluate the accuracy of the measurement with microchip epidural system.Methods:
Twenty-seven pigs with were studied, under generaI anesthesia, properly assisted with ventilation and hemodynamic monitoring. During the experiment, we have simulated frontal intracerebral expansive process. The experiment consisted of three groups (A, B and C) with intracranial hypertension generated with the simulation of an intracerebral hematoma. The two systems were compared andstudied as the correlation of the measured values.Results:
The average behavior of the increased intracranial pressure (ICP) over the time points are statistically different between groups (p < 0.001). The simulation of rebleeding resulted in a significant increase in ICP (p < 0.001). Evaluating the overall comparative accuracy there was an intraclass correlation coefficient of 0.8.Conclusion:
The model of intracranial hypertension balloon in pigs is feasible and reliable in generating intracranial hypertension. The system for measuring intracranial epidural pressure has a high correlation coefficient with the system parenchymal gauging the overall evaluation.
Texto completo:
Disponible
Índice:
LILACS (Américas)
Asunto principal:
Presión Intracraneal
/
Hemorragia Cerebral
/
Hipertensión Intracraneal
Límite:
Animales
Idioma:
Portugués
Revista:
Arq. bras. neurocir
Asunto de la revista:
Cirurgia
/
Neurocirugía
Año:
2014
Tipo del documento:
Artículo
País de afiliación:
Brasil
Institución/País de afiliación:
Faculdade de Medicina/BR
/
Medicina da Universidade de São Paulo/BR
/
Santa Casa de Passos/BR
/
Universidade de São Paulo/BR
Similares
MEDLINE
...
LILACS
LIS