Rastreio e manejo de pacientes com cabos-eletrodos de cardiodesfibriladores Riata com condutores externalizados: revisão da literatura / Screening and management of patients with Riata implantable cardioverter-defibrillator eads with externalized conductors: literature review
Ladeira, Fernanda Brasiliense; Camanho, Luiz Eduardo Montenegro; Slater, Charles; Inácio Junior, Luiz Antonio de Oliveira; Dias, Lucas Carvalho; Saad, Eduardo Benchimol.
  • Ladeira, Fernanda Brasiliense; Hospital Pró-Cardíaco. Setor de Arritmia Invasiva. Rio de Janeiro - RJ. BR
  • Camanho, Luiz Eduardo Montenegro; Hospital Pró-Cardíaco. Serviço de Arritmia Invasiva. Rio de Janeiro - RJ. BR
  • Slater, Charles; Hospital Pró-Cardíaco. Serviço de Arritmia Invasiva. Rio de Janeiro - RJ. BR
  • Inácio Junior, Luiz Antonio de Oliveira; Hospital Pró-Cardíaco. Serviço de Arritmia Invasiva. Rio de Janeiro - RJ. BR
  • Dias, Lucas Carvalho; Hospital Pró-Cardíaco. Serviço de Arritmia Invasiva. Rio de Janeiro - RJ. BR
  • Saad, Eduardo Benchimol; Hospital Pró-Cardíaco. Serviço de Arritmia Invasiva. Rio de Janeiro - RJ. BR
RELAMPA, Rev. Lat.-Am. Marcapasso Arritm ; 28(4): 147-154, out.-dez. 2015. ilus, tab
Artículo en Portugués | LILACS | ID: lil-788751
RESUMO
Os cabos-eletrodos de cardiodesfibriladores com revestimento de silicone Riata e Riata ST foramextensamente implantados entre os anos de 2003 e 2010. Sua distribuição, porém, foi interrompida após aobservação de exteriorização dos condutores por falha do isolante em diversos casos de mau funcionamento.Esses cabos-eletrodos foram classificados como classe I de recall pelo Food and Drug Administration. Diversosestudos foram realizados desde então para se determinar a incidência e a prevalência de extrusão dos condutores epara avaliar sua correlação com falha elétrica. Também tem sido estudado o melhor método de rastreio e como seconduzir os pacientes portadores desses cabos-eletrodos.
ABSTRACT
The cardioverter-defibrillator leads with Riata silicone coating and Riata ST were widely deployed between 2003 and 2010. Its distribution, however, was discontinued after observation of externalization of conductors due to insulation failure in several cases of malfunction. These leads were classified as Class I Recall by the Food and Drug Administration. Several studies have been performed since then to determine the incidence and prevalence of extrusion of the conductors and to evaluate their correlation with electrical failure. The best screening method and how to conduct patients with these leads have also been investigated.
Asunto(s)

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Arritmias Cardíacas / Desfibriladores Implantables / Electrodos Implantados / Falla de Equipo Tipo de estudio: Estudio diagnóstico / Factores de riesgo / Estudio de tamizaje Límite: Humanos Idioma: Portugués Revista: RELAMPA, Rev. Lat.-Am. Marcapasso Arritm Asunto de la revista: Cardiología Año: 2015 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Hospital Pró-Cardíaco/BR

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Arritmias Cardíacas / Desfibriladores Implantables / Electrodos Implantados / Falla de Equipo Tipo de estudio: Estudio diagnóstico / Factores de riesgo / Estudio de tamizaje Límite: Humanos Idioma: Portugués Revista: RELAMPA, Rev. Lat.-Am. Marcapasso Arritm Asunto de la revista: Cardiología Año: 2015 Tipo del documento: Artículo País de afiliación: Brasil Institución/País de afiliación: Hospital Pró-Cardíaco/BR