Síndrome de marcapasos por selección inadecuada del modo de estimulación / Pacemaker syndrome due to inadequate selection in the stimulation way
Angulo Elers, Carlos Manuel; Reyes Sánchez, Raúl Ernesto; García Salas, Enrique Antonio.
  • Angulo Elers, Carlos Manuel; Universidad de Ciencias Médicas. Hospital Provincial Docente Clinicoquirúrgico Saturnino Lora Torres. Santiago de Cuba. CU
  • Reyes Sánchez, Raúl Ernesto; Universidad de Ciencias Médicas. Hospital Provincial Docente Clinicoquirúrgico Saturnino Lora Torres. Santiago de Cuba. CU
  • García Salas, Enrique Antonio; Universidad de Ciencias Médicas. Hospital Provincial Docente Clinicoquirúrgico Saturnino Lora Torres. Santiago de Cuba. CU
Medisan ; 20(10)oct. 2016. ilus
Artículo en Español | LILACS, CUMED | ID: lil-797510
RESUMEN
Se presenta el caso clínico de un paciente de 52 años de edad que, a causa de una disfunción sintomática del nódulo sinusal, portaba un marcapasos permanente unicameral ventricular desde hacía 9 años, al cual se le había sustituido el generador por agotamiento de la batería. Un mes después del cambio el paciente acudió a la consulta especializada de Arritmias y Marcapasos en el Hospital Provincial Docente Clinicoquirúrgico "Saturnino Lora Torres" de Santiago de Cuba, por presentar síntomas de cansancio fácil, disnea, mareos y síncopes; se le indicaron electrocardiogramas de superficie que mostraron una conducción retrógrada ventriculoauricular y pseudofusiones. Teniendo en cuenta las manifestaciones clínicas y los hallazgos electrocardiográficos, se diagnosticó un síndrome de marcapasos y se decidió implantar un electrodo auricular y variar la modalidad de estimulación a la de doble cámara, como terapéutica efectiva para ello. Los síntomas desaparecieron y el paciente evolucionó favorablemente hasta su egreso de la institución hospitalaria.
ABSTRACT
The case report of a 52 years patient is presented that had a ventricular unicameral permanent pacemaker for 9 years, due to a symptomatic dysfunction of the synusal nodule. The pacemaker had dead battery and the generator was substituted. A month after the change the patient went to the Arrhytmias and Pacemaker specialized service at "Saturnino Lora Torres" Teaching Clinical Surgical Provincial Hospital in Santiago de Cuba, due to symptoms of easy tireness, dyspnea, dizziness and fainting fits; surface electrocardiograms were indicated that showed a ventriculoauricular retrograde conduction and pseudofusions. Taking into account the clinical features and the electrocardiographic findings, a pacemaker syndrome was diagnosed and it was decided to implant an auricular electrode and vary the stimulation modality to that of double camera, as effective therapy for it. The symptoms disappeared and the patient had a favorable clinical course when he was discharged from the hospital institution.
Asunto(s)

Texto completo: Disponible Índice: LILACS (Américas) Asunto principal: Marcapaso Artificial Idioma: Español Revista: Medisan Asunto de la revista: Medicina Año: 2016 Tipo del documento: Artículo País de afiliación: Cuba Institución/País de afiliación: Universidad de Ciencias Médicas/CU