Your browser doesn't support javascript.
loading
Comparison of Forehead Infrared Thermometry with Axillary Digital Thermometry in Neonates.
Indian Pediatr ; 2013 December; 50(12): 1153-1154
Artículo en Inglés | IMSEAR | ID: sea-170098
ABSTRACT
Fluctuations in core body temperature beyond a narrow range are of concern as they indicate changed homeostasis. Thermoregulation remains a key aspect of neonatal care. New ‘non touchmethod of measuring temperature may reduce infection rate and discomfort of neonates. Body temperature of neonates admitted in Neonatal Intensive Care Unit was measured using axillary digital thermometer and a handheld infrared non touch thermometer. The two methods did not agree well (mean difference = -0.5, 95% limits of agreement [-2.3, 1.2]). The agreement was similar with a negligible difference when patients in open care warmers were excluded (mean difference = -0.6, 95% limits of agreement [-2.3, 1.1]). As the accuracy is unsatisfactory, the said technology needs further improvement before it can be adapted for routine patient care.

Texto completo: Disponible Índice: IMSEAR (Asia Sudoriental) Idioma: Inglés Revista: Indian Pediatr Año: 2013 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: IMSEAR (Asia Sudoriental) Idioma: Inglés Revista: Indian Pediatr Año: 2013 Tipo del documento: Artículo