A modified formula for defining tissue phantom ratio of photon beams.
Bangladesh Med Res Counc Bull
;
2007 Dec; 33(3): 92-7
Artículo
en Inglés
| IMSEAR
| ID: sea-258
ABSTRACT
Tissue phantom ratio (TPR), for square fields of various dimensions has been determined at varying depths in water. The dose in water has been measured at a fixed source-to-surface distance (SSD) of 100 cm and reference depth of 5 cm for 6 MV photon beam of Siemens Linear Accelerator Primus 11 in German Cancer Research Center (DKFZ), Heidelberg, Germany. A modified formula has been developed to calculate the TPR value for isocentric treatment. The present article describes the conversion of the measured data values into a comprehensive and consistent data set by the modified formula, that gives the TPR from Percentage Depth Dose (PDD) with depth as a function of field sizes from 10 mm x 10 mm upto 300 mm x 300 mm) and depth (from 0 mm to 300 mm).
Texto completo:
Disponible
Índice:
IMSEAR (Asia Sudoriental)
Asunto principal:
Radiometría
/
Humanos
/
Absorciometría de Fotón
/
Fotones
/
Fantasmas de Imagen
/
Modelos Teóricos
Idioma:
Inglés
Revista:
Bangladesh Med Res Counc Bull
Año:
2007
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS