Your browser doesn't support javascript.
loading
Development of an amperometric hypoxanthine biosensor for determination of hypoxanthine in fish and meat tissue.
Indian J Exp Biol ; 2005 Jul; 43(7): 646-53
Artículo en Inglés | IMSEAR | ID: sea-58028
ABSTRACT
A hypoxanthine (Hx) biosensor based on immobilized xanthine oxidase (XO) as the bio-component was developed and studied for the rapid analysis of fish (sweet water and marine) and goat meat samples. The biosensor was standardized for the determination of Hx in the range of 0.05 to 2 mM. Crosslinking with glutaraldehyde in presence of BSA as a spacer molecule was used for the method of immobilization. One layer of gelatin (10%) was applied over the immobilized enzyme layer to reduce the leaching out of enzyme from the membrane (cellulose acetate) matrix. The optimum pH of the immobilized system was determined to be 8.5 at 25 degrees C instead 7.0-7.2 for free enzyme system. Km and Vmax values were determined for the immobilized system. The developed sensor was applied to determine the amount of Hx present in fish and meat over a period of time. The stability of the enzyme immobilized membrane was also tested over a period of 30 days.
Asunto(s)
Texto completo: Disponible Índice: IMSEAR (Asia Sudoriental) Asunto principal: Xantina Oxidasa / Bovinos / Técnicas Biosensibles / Hipoxantina / Enzimas Inmovilizadas / Peces / Análisis de los Alimentos / Animales / Carne Idioma: Inglés Revista: Indian J Exp Biol Año: 2005 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: IMSEAR (Asia Sudoriental) Asunto principal: Xantina Oxidasa / Bovinos / Técnicas Biosensibles / Hipoxantina / Enzimas Inmovilizadas / Peces / Análisis de los Alimentos / Animales / Carne Idioma: Inglés Revista: Indian J Exp Biol Año: 2005 Tipo del documento: Artículo