An audit of phototherapy units.
Indian J Pediatr
;
2000 Dec; 67(12): 883-4
Artículo
en Inglés
| IMSEAR
| ID: sea-80368
ABSTRACT
Audit of phototherapy aims to examine their efficacy and functioning. Twenty-four centres providing neonatal care were visited by a team of two persons. Data form were completed with information regarding age of phototherapy units, its buildup in relation to number of tubelights, type of light and irradiance provided. Total of 58 units examined had a wide variety in relation to their build. There were only 21 units (36.2%) in which all the lights were in working order. Only five of the units (8.6%) had the recommended special blue lights. Only 18 of the units (31%) provided an acceptable level of irradiance. Phototherapy demonstrates a dose response relationship. By not providing optimum irradiance, the efficacy is compromised. This prolongs hospital stay and treatment costs. Acceptable standards should be insisted upon in purchase and maintenance of medical equipments.