Your browser doesn't support javascript.
loading
Impedance drop determines ablation lesion volume at the same level of ablation index
International Journal of Arrhythmia ; : 10-2023.
Artículo en Inglés | WPRIM | ID: wpr-1000500
ABSTRACT
New parameters such as ablation index (AI) have been developed to create reliable ablation lesions. This study was performed to evaluate whether RF energy delivery with the same ablation index creates the similar ablation lesion volume. Ablation lesions were created in 5 pig hearts at ex-vivo state. Ablation was performed using an external-irri‑ gation contact-force sensing catheter on the epicardial side of the left ventricle with 90-degree of angle. RF ablation time was adjusted for targeting AI 600 at 8 different conditions. Lesion volume created with 0–5 g of contact force at 20 W was significantly lower than that of 11–20 g at 40W despite of the same AI (125 ± 76.2 vs. 272 ± 49.5 ­mm 3power for the ablation of exvivo swine left ventricle, and high-quality lesions could be expected when the impedance drop is satisfactory even though the same level of AI is applied during ablation.
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Inglés Revista: International Journal of Arrhythmia Año: 2023 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Inglés Revista: International Journal of Arrhythmia Año: 2023 Tipo del documento: Artículo