Your browser doesn't support javascript.
loading
Radiation environment monitoring and analysis of exempted X-ray devices / 中国辐射卫生
Chinese Journal of Radiological Health ; (6): 511-516, 2023.
Artículo en Chino | WPRIM | ID: wpr-1003555
ABSTRACT
Objective To propose specific suggestions for exemption management of X-ray devices applying for exemption management. Methods The weight analysis of influence of shielding thickness and tube voltage on the radiation dose leakage of low-energy X-ray devices was carried out through test analysis, and the radiation status of 50 X-ray devices applied for end-user exemption in Shenzhen, China was investigated. Results There was no significant correlation between the ambient dose equivalent rate and the tube voltage and tube current of the X-ray device. The ambient dose equivalent rate of 50 X-ray devices was 0.16-1.20 μSv/h. Among them, 8 X-ray devices had radiation dose leakage, that is, the ambient dose equivalent rate was greater than 0.36 μSv/h; the ambient dose equivalent rate of one X-ray device was 1.2 μSv/h, exceeding exemption management requirements. Conclusion Applications for exemption management and filing of X-ray devices should be treated differently, and their own shielding effect and inherent safety have a greater impact on radiation safety, which should be focused on.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiological Health Año: 2023 Tipo del documento: Artículo