Measurement of 239Pu in fecal samples based on inductively coupled plasma-mass spectrometry / 中国辐射卫生
Chinese Journal of Radiological Health
;
(6): 632-635, 2023.
Artículo
en Chino
| WPRIM
| ID: wpr-1006318
ABSTRACT
Objective To establish a method for measurement of 239Pu in fecal samples based on inductively coupled plasma-mass spectrometry (ICP-MS), and to provide a novel method for assessing the internal exposure of workers. Methods Fecal samples were collected from workers and labeled. The samples were pretreated with carbonization ashing and microwave digestion devices, purified on TEVA resin, and measured using ICP-MS. Results The detection limit of 239Pu in fecal samples based on ICP-MS was 1.91 × 10−4 Bq. Conclusion In the routine monitoring of class S substances characterized by a 5 μm aerodynamic diameter during 12 months, the committed effective dose corresponding to the detection limit is 0.17 mSv. This value meets the requirements of relevant national standards and ICP-MS can be used as a novel means for accurate evaluation of internal exposure for workers.