Your browser doesn't support javascript.
loading
Conditions for the determination of intrinsic uniformity and intrinsic spatial linearity of SPECT / 中国辐射卫生
Chinese Journal of Radiological Health ; (6): 661-667, 2023.
Artículo en Chino | WPRIM | ID: wpr-1006324
ABSTRACT
Objective To address the absence of matrix specified for the determination of intrinsic uniformity in the current standard, and to investigate the effect of source distances on intrinsic spatial linearity, the intrinsic uniformity and intrinsic spatial linearity of 16 probes in eight SPECT devices were measured and analyzed with different matrices and source distances, in order to determine the optimal measurement conditions. Methods According to the standard Specification for Testing of Quality Control in Gamma Cameras and Single Photon Emission Computed Tomograph (SPECT) (WS 523—2019), the intrinsic uniformity was measured using 64 × 64 and 256 × 256 matrices and the intrinsic spatial linearity was measured using of 1.7 and 3 m source distances. Results When intrinsic uniformity was measured with the 64 × 64 matrix, more than 50% of the probes showed lower values. When intrinsic spatial linearity was measured with the 3 m source distance, more probes showed lower values. Conclusion The 64 × 64 matrix is recommended for the determination of intrinsic uniformity and a source distance of >5 FOV is recommended for the measurement of intrinsic spatial linearity.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiological Health Año: 2023 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiological Health Año: 2023 Tipo del documento: Artículo