MR Imaging of Slow-flow Using a Flow Phantom
Journal of the Korean Society of Magnetic Resonance in Medicine
;
: 116-122, 2001.
Artículo
en Inglés
| WPRIM
| ID: wpr-10129
ABSTRACT
Purpose:
To find sensitivity of MRI imaging methods to slow flow phantom study was performed with conventional Spin-Echo, gradient echo based Phase Contrast, fast GLASS, and heavily T2-weighted Fast Spin Echo pulse sequences. MATERIALS ANDMETHODS:
A siphon driven flow phantom was constructed with a ventriculo-peritoneal shunt catheter and a GE phantom to achieve continuous variable flow. Four different pulse sequences including Spin-Echo, Phase Contrast, GRASS and Heavily T2-weighted Fast Spin Echo were evaluated to depict slow flow in the range from 0.08 ml/min to 1.7 ml/min and to compare signal intensities between static fluid and flowing fluid.RESULTS:
In the slow flow above 0.17 ml/min conventional Spin-Echo showed superior apparent contrast between static and flowing fluid while GRASS was more sensitive to the very slow flow below 0.17 ml/min. It was not accurate to calculate flow and velocity below 0.1 ml/min with a modified PC imaging.CONCLUSION:
Four different MR pulse sequences demonstrated different sensitivity to the range of slow flow from 0.08 ml/min to 1.7 ml/min. This finding may be clinically useful to measure CSF shunt flow or detecting CSF collection and thrombosis.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Trombosis
/
Imagen por Resonancia Magnética
/
Derivación Ventriculoperitoneal
/
Catéteres
/
Vidrio
/
Poaceae
Idioma:
Inglés
Revista:
Journal of the Korean Society of Magnetic Resonance in Medicine
Año:
2001
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS