Your browser doesn't support javascript.
loading
Determination of four organotin compounds in human whole blood by high performance liquid chromatography-inductively coupled plasma-mass spectrometry / 中国职业医学
China Occupational Medicine ; (6): 712-716, 2023.
Artículo en Chino | WPRIM | ID: wpr-1013312
ABSTRACT
{L-End}Objective To establish a method for the simultaneous determination of dimethyltin (DMT), trimethyltin (TMT), diethyltin (DET), and triethyltin (TET) in human whole blood using high performance liquid chromatography-inductively coupled plasma-mass spectrometry (ICP-MS). {L-End}Methods The 1.0 mL of blood was added with 4.0 mL 65% aqueous solution (containing 6% acetic acid), extracted and separated by C4 column (150 mm×3 mm×3 μm) using a mobile phase of methanol and 4% acetic acid aqueous solution (containing 0.25 mmol/L tropolone) at a volume ratio of 35∶65, and detected by ICP-MS. {L-End}Results The linear range of DMT, TMT, DET, and TET was 30.60-550.80, 29.00-522.00, 46.10-829.80, and 34.05-612.90 μg/L, respectively. All correlation coefficients were 0.999. The detection limit of DMT, TMT, DET and TET was 21.40, 20.30, 32.27 and 23.80 μg/L, respectively. The recovery rate was 81.9%-104.9%. The within-run and between-run relative standard deviation was 1.6%-6.9% and 0.1%-10.0%, respectively. The samples can be stored at -20 ℃ and 4 ℃ for at least three days. {L-End}Conclusion This method can be used for trace analysis of DMT, TMT, DET, and TET in whole blood.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: China Occupational Medicine Año: 2023 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: China Occupational Medicine Año: 2023 Tipo del documento: Artículo