We developed a simple, sensitive, and effective ultra-performance liquid chromatography/tandem mass spectrometry (HPLC-MS/MS) method with an electrospray ionization (ESI) interface in multiple reaction monitoring (MRM) and positive ion modes to determine diazepam concentrations in human plasma using voriconazole as an internal standard (IS). Diazepam and IS were detected at transition 285.2→193.1 and 350.2→127.1, respectively. After liquid-liquid extraction (LLE) using 1.2 ml of ethyl acetaten-hexane (8020, v/v), diazepam and IS were eluted on a Phenomenex Cadenza CD-C18 column (150 × 3.0 mm, 3 µm) with an isocratic mobile phase (10 mM ammonium acetate in watermethanol [595, v/v]) at a flow rate of 0.4 mL/min. The peak retention time was 2.32 min for diazepam and 2.01 min for IS, respectively. The lower limit of quantitation (LLOQ) was 0.5 ng/mL (S/N > 10) using 50 µL of plasma, and no interferences were observed in chromatograms. Our analytical method was fully validated and successfully applied to a bioequivalence study of two formulations of diazepam in healthy Korean volunteers.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Plasma
/
Espectrometría de Masas
/
Voluntarios
/
Equivalencia Terapéutica
/
Diazepam
/
Extracción Líquido-Líquido
/
Compuestos de Amonio
/
Voriconazol
/
Métodos
Tipo de estudio:
Estudio diagnóstico
Límite:
Humanos
Idioma:
Inglés
Revista:
Translational and Clinical Pharmacology
Año:
2017
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS