ABSTRACT

PURPOSE:

To validate the Dinamap ProCare 200 blood pressure (BP) monitor against a mercury sphygmomanometer in children 7 to 18 years old in accordance with the 2010 International Protocol of European Society of Hypertension (ESH-IP2) and the British Hypertension Society (BHS) protocol.

METHODS:

Forty-five children were recruited for the study. A validation procedure was performed following the protocol based on the ESH-IP2 and BHS protocols for children and adolescents. Each subject underwent 7 sequential BP measurements alternatively with a mercury sphygmomanometer and the test device by trained nurses. The results were analyzed according to the validation criteria of ESH-IP2.

RESULTS:

The mean (+/-SD) difference in the absolute BP values between test device and mercury sphygmomanometer readings was 1.85+/-1.65 mmHg for systolic BP (SBP) and 4.41+/-3.53 mmHg for diastolic BP (DBP). These results fulfilled the Association for the Advancement of Medical Instrumentation criterion of a mean+/-SD below 5+/-8 mmHg for both SBP and DBP. The percentages of test device-observer mercury sphygmomanometer BP differences within 5, 10, and 15 mmHg were 96%, 100%, and 100% for SBP, and 69%, 92%, and 100% for DBP, respectively, in the part 1 analysis; both SBP and DBP passed the part 1 criteria. In the part 2 analysis, SBP passed the criteria but DBP failed.

CONCLUSION:

Although the Dinamap ProCare 200 BP monitor failed an adapted ESH-IP2, SBP passed. When comparing BP readings measured by oscillometers and mercury sphygmomanometers, one has to consider the differences between them, particularly in DBP, because DBP can be underestimated.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Compuestos Organotiofosforados / Brazo / Lectura / Presión Sanguínea / Monitores de Presión Sanguínea / Esfigmomanometros / Hipertensión Límite: Adolescente / Niño / Humanos Idioma: Inglés Revista: Korean Journal of Pediatrics Año: 2011 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Compuestos Organotiofosforados / Brazo / Lectura / Presión Sanguínea / Monitores de Presión Sanguínea / Esfigmomanometros / Hipertensión Límite: Adolescente / Niño / Humanos Idioma: Inglés Revista: Korean Journal of Pediatrics Año: 2011 Tipo del documento: Artículo