Effects of an Extra-corporeal Life Support System Using a Dual Pulsatile Pump
Journal of the Korean Society of Emergency Medicine
;
: 489-496, 2002.
Artículo
en Coreano
| WPRIM
| ID: wpr-147257
ABSTRACT
PURPOSE:
The purpose of this study was to observe and compare the changes in the patterns of hemodynamic and blood profiles on the circuit of Extra-corporeal life support system (ECLS) by using a dual pulsatile pump (T-PLS).METHODS:
An acute heart failure model using partial pulmonary artery banding was constructed in 12 piglets (20 -25kg). The animals were divided into centrifugal (n=6) and dual pulsatile pump (n=6) group. Each animal was placed on an ECLS system with a membrane oxygenator bypassing the right atrium and the aorta for 2 hours under general anesthesia. The parameters mainly observed were intra-circuit pressure changes, arterial pulsatility (pulse pressure), plasma free hemoglobin, hemodynamic changes, and other blood profiles. The parameters obtained just prior to the bypass were compared with the corresponding parameters obtained two hours after the bypass.RESULTS:
Before bypass, the parameters were statistically the same between the groups. Two hours after the bypass, no significant differences were observed between the groups in ABGA, VBGA, AST/ALT, BUN/Cr, and electrolytes; the plasma free hemoglobin was 14.8+/-4.7 g/dl in the dual pulsatile group and 19.1+/-9.1 g/dl in the centrifugal group (p=NS). The pulse pressure was higher in the dual pulsatile pump than in the centrifugal pump group (35+/-8 vs. 11+/-7 mmHg, p=0.0253 mmHg). The highest circuit pressure was generated at the inlet of the membrane oxygenator and was higher in the dual pulsatile group than in the centrifugal group (173+/-12 mmHg vs. 222+/-8 mmHg, p=0.0000).CONCLUSION:
The results demonstrate that a dual pulsating mechanism lessens blood cell trauma while providing physiologic pulsatile blood flow. The ECLS system using a dual pulsatile pump (T-PLS) can be useed as an effective and safe driving motor for an ECLS.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Aorta
/
Oxigenadores de Membrana
/
Plasma
/
Arteria Pulmonar
/
Células Sanguíneas
/
Presión Sanguínea
/
Reanimación Cardiopulmonar
/
Bahías
/
Electrólitos
/
Atrios Cardíacos
Límite:
Animales
Idioma:
Coreano
Revista:
Journal of the Korean Society of Emergency Medicine
Año:
2002
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS