Comparison of Corneal Thickness Measurements with Noncontact Specular Microscope and Ultrasonic Pachymeter
Journal of the Korean Ophthalmological Society ; : 1060-1065, 2004.
Artículo en Coreano | WPRIM | ID: wpr-15050
ABSTRACT

PURPOSE:

To evaluate the corneal thickness values before and after refractive keratoplasty with the Topcon SP-2000 noncontact specular microscope(Topcon American Corp.) and the common standard ultrasonic pachymeter.

METHODS:

Corneal thickness was measured in 40 eyes of 20 patients, 5 times each before and after keratoplasty, with the Topcon SP-2000P and an ultrasonic pachymeter (DGH 1000, DGH Technology Inc.). The main outcome measures were accuracy and reproducibility as assessed with coefficient of variation (CV) of corneal thickness measurements.

RESULTS:

The reproducibility of corneal thickness measurements was better with the ultrasonic pachymeter (CV 0.36%) than with the Topcon SP-2000 (CV 1.23%). Mean corneal thickness measured by the Topcon SP-2000P before and after keratoplasty was significantly less than the mean values obtained with the ultrasonic pachymeter (14.4 micro meter; P<.0001, 5.6 micro meter; P<.01; respectively). Compared with theoretical ablated corneal thickness, ablated corneal thickness measured with the Topcon SP-2000P was significantly less (P<.01) than values obtained with the ultrasonic pachymeter (P=0.07)

CONCLUSIONS:

The noncontact Topcon specular microscope provides measurements of corneal thickness that are significantly less than those of the ultrasonic pachymeter, both before and after keratoplasty. These results indicate that these 2 techniques are not clinically comparable, thereby precluding interchangeable use of their data in planning or assessing keratoplasty.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Ultrasonido / Trasplante de Córnea / Evaluación de Resultado en la Atención de Salud / Paquimetría Corneal Límite: Humanos Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Año: 2004 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Ultrasonido / Trasplante de Córnea / Evaluación de Resultado en la Atención de Salud / Paquimetría Corneal Límite: Humanos Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Año: 2004 Tipo del documento: Artículo