Left Ventricular Ejection Fraetion Determmed by Cated Tl-201 Perfusion SPECT and Quantitative Software / 대한핵의학회잡지
Korean Journal of Nuclear Medicine
; : 222-227, 2000.
Article
en Ko
| WPRIM
| ID: wpr-151753
Biblioteca responsable:
WPRO
ABSTRACT
PURPOSE: We compared estimates of ejection fraction (EF) determined by gated Tl-201 perfusion SPECT (g-TI-SPECT) with those by gated blood pool (GBP) scan. MATERIALS AND METHODS: Eighteen subjects underwent g-TI-SPECT and GBP scan. After reconstruction of g-TI-SPECT, we measured EF with Cedars software. The comparison of the EF with g-TI-SPECT and GHP scan was assessed by correlation analysis and Bland Altman plot. RESULTS: The estimates of EF were significantly different (p<0.05) with g-TI-SPECT (40%+/-14%) and GBP scan (43%+/-14%). There was an excellent correlation of EF between e-TI-SPECT and GBP scan (r=0.94, p<0.001). The mean difference of EF between GRP scan and g-TI- SPECT was +3.2%, Ninety-five percent limits of agreement were +9,8%. EF between g-TI-SPECT and GBP scan were in poor agreement. CONCLUSION: The estimates of EF by g-TI-SPECT was well correlated with those by GBP scan. However, EF of g-TI-SPECT doesn't agree with EF of GBP scan. EF of g-TI-SPECT cant be used interchangeably with EF of GBP scan.
Palabras clave
Texto completo:
1
Índice:
WPRIM
Asunto principal:
Perfusión
/
Tomografía Computarizada de Emisión de Fotón Único
Idioma:
Ko
Revista:
Korean Journal of Nuclear Medicine
Año:
2000
Tipo del documento:
Article