BACKGROUND:
We evaluated the analytical and clinical performances of the SD BIOLINE Rota/Adeno Rapid kit (SD Rota/Adeno Rapid; Standard Diagnostics, Inc., Korea), an immunochromatographic assay (ICA), for the simultaneous detection of rotaviruses and adenoviruses in human stool samples.METHODS:
We tested 400 clinical stool samples from patients with acute gastroenteritis and compared the ICA results with the results obtained by using ELISA, enzyme-linked fluorescent assays (ELFA), PCR, and multiplex reverse transcription-PCR (mRT-PCR). To assess the analytical performance of the SD BIOLINE Rota/Adeno Rapid kit, we determined its detection limit, reproducibility, cross-reactivity, and analytical reactivity for adenovirus subtypes, and performed interference studies.RESULTS:
The overall agreement rates among the tested methods were 91.5% for rotavirus and 85.5% for adenovirus. On the basis of mRT-PCR, the overall agreement, positive agreement, and negative agreement rates of the ICA were 95.6%, 100%, and 94.9% for rotavirus, and 94.0%, 71.4%, and 94.8% for adenovirus, respectively. Using the ICA, we detected all the subtypes of adenovirus tested, but the analytical reactivities for adenovirus subtypes were different between the 4 adenovirus detection methods. The high reproducibility was confirmed, and no cross-reactivity or interference was detected.CONCLUSIONS:
The SD BIOLINE Rota/Adeno Rapid kit showed acceptable analytical and clinical performances. However, interpretation of adenovirus positive/negative result should be cautious because of different detectability for adenovirus subtypes among adenovirus detection methods.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Juego de Reactivos para Diagnóstico
/
ADN Viral
/
ARN Viral
/
Ensayo de Inmunoadsorción Enzimática
/
Adenoviridae
/
Enfermedad Aguda
/
Reproducibilidad de los Resultados
/
Cromatografía de Afinidad
/
Rotavirus
/
Reacción en Cadena de la Polimerasa de Transcriptasa Inversa
Tipo de estudio:
Estudio diagnóstico
Límite:
Humanos
Idioma:
Inglés
Revista:
Annals of Laboratory Medicine
Año:
2014
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS