Evaluation of the Performance of a Micromethod for Measuring Urinary Iodine by Using Six Sigma Quality Metrics
Annals of Laboratory Medicine
;
: 319-325, 2013.
Artículo
en Inglés
| WPRIM
| ID: wpr-178349
ABSTRACT
BACKGROUND:
The urinary iodine micromethod (UIMM) is a modification of the conventional method and its performance needs evaluation.METHODS:
UIMM performance was evaluated using the method validation and 2008 Iodine Deficiency Disorders survey data obtained from four urinary iodine (UI) laboratories. Method acceptability tests and Sigma quality metrics were determined using total allowable errors (TEas) set by two external quality assurance (EQA) providers.RESULTS:
UIMM obeyed various method acceptability test criteria with some discrepancies at low concentrations. Method validation data calculated against the UI Quality Program (TUIQP) TEas showed that the Sigma metrics were at 2.75, 1.80, and 3.80 for 51+/-15.50 microg/L, 108+/-32.40 microg/L, and 149+/-38.60 microg/L UI, respectively. External quality control (EQC) data showed that the performance of the laboratories was within Sigma metrics of 0.85-1.12, 1.57-4.36, and 1.46-4.98 at 46.91+/-7.05 microg/L, 135.14+/-13.53 microg/L, and 238.58+/-17.90 microg/L, respectively. No laboratory showed a calculated total error (TEcalc)CONCLUSIONS:
UIMM showed unacceptable performance for the iodine deficiency levels and variable performance at other concentrations according to different TEas.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Control de Calidad
/
Espectrofotometría
/
Urinálisis
/
Yodo
/
Laboratorios
Límite:
Humanos
Idioma:
Inglés
Revista:
Annals of Laboratory Medicine
Año:
2013
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS