Your browser doesn't support javascript.
loading
Comparative study of abutment screw loosening with or without adhesive material
The Journal of Advanced Prosthodontics ; : 99-103, 2017.
Artículo en Inglés | WPRIM | ID: wpr-179524
ABSTRACT

PURPOSE:

The purpose of this study was to achieve more retention and stability and to delay or prevent screw loosening. MATERIALS AND

METHODS:

Twenty implants (Implantium 3.4 mm, Dentium, Seoul, Korea) were divided into 2 groups (n = 20). In the first group, an adhesive material was applied around the screw of the abutments (test group). In the second group, the screws are soaked in saliva (control group). All the screws were torqued under 30 N/cm, Then, the samples were gone through a cyclic fatigue loading process. After cyclic loading, we detorqued screws and calculated detorque value.

RESULTS:

In comparison with the control group, all the implant screws in the test group were smeared with the adhesive material, showing significant higher detorque value.

CONCLUSION:

There are significantly higher detorque values in the group with adhesive. It is recommended to make biocompatible adhesive to reduce screw loosening.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Saliva / Implantes Dentales / Adhesivos / Fatiga / Diseño de Implante Dental-Pilar / Seúl País/Región como asunto: Asia Idioma: Inglés Revista: The Journal of Advanced Prosthodontics Año: 2017 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Saliva / Implantes Dentales / Adhesivos / Fatiga / Diseño de Implante Dental-Pilar / Seúl País/Región como asunto: Asia Idioma: Inglés Revista: The Journal of Advanced Prosthodontics Año: 2017 Tipo del documento: Artículo