Evaluation of a Novel Array-Based Toxoplasma, Rubella, Cytomegalovirus, and Herpes Simplex Virus IgG Enzyme Linked Immunosorbent Assay and Its Comparison with Virion/Serion Enzyme Linked Immunosorbent Assays
Annals of Laboratory Medicine
;
: 38-42, 2014.
Artículo
en Inglés
| WPRIM
| ID: wpr-193130
ABSTRACT
BACKGROUND:
The dramatic increase in use of the IgG test for toxoplasma, rubella, cytomegalovirus (CMV), and herpes simplex virus (HSV) [TORCH] has led to the requirement for a high-efficiency method that can be used in the clinical laboratory. This study aimed to compare the results of BGI-Array ELISA TORCH IgG (BGI-GBI, China) screening method to those of Virion/Serion TORCH IgG ELISA (Virion/Serion, Germany).METHODS:
Serum specimens (n=400) submitted for routine IgG testing by Virion/Serion ELISA were also tested using the BGI-Array ELISA method. The agreements of these two kinds of method were analyzed by kappa-coefficients calculation.RESULTS:
Following repeat testing, the BGI-Array ELISA TORCH IgG assays demonstrated agreements of 99.5% (398/400 specimens), 98% (392/400 specimens), 99% (396/400 specimens), and 99.5% (398/400 specimens), respectively. The BGI-Array ELISA IgG assays provided results comparable to Virion/Serion ELISA results, with kappa-coefficients showing near-perfect agreement for the HSV (kappa=0.87), rubella (kappa=0.92) and CMV (kappa=0.93) and substantial agreement for the toxoplasma (kappa=0.80) IgG assays. The use of the BGI-Array ELISA TORCH IgG assays could reduce the turnaround time (1.5 hr vs. 5 hr by Virion/Serion ELISA for 100 specimens) and were easy to use.CONCLUSIONS:
BGI-Array ELISA TORCH IgG shows a good agreement with Virion/Serion ELISA methods and is suitable for clinical application.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Infecciones por Protozoos
/
Juego de Reactivos para Diagnóstico
/
Virus de la Rubéola
/
Toxoplasma
/
Virión
/
Virosis
/
Inmunoglobulina G
/
Ensayo de Inmunoadsorción Enzimática
/
Sensibilidad y Especificidad
/
Simplexvirus
Tipo de estudio:
Estudio diagnóstico
Límite:
Humanos
Idioma:
Inglés
Revista:
Annals of Laboratory Medicine
Año:
2014
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS