The Breakage of Epidural Catheter during Continuous Epidural Analgesia for Cancer Pain: A case report / 대한마취과학회지
Korean Journal of Anesthesiology
;
: 513-515, 2007.
Artículo
en Coreano
| WPRIM
| ID: wpr-193259
ABSTRACT
During pain control by continuous epidural analgesia, the breakage of epidural catheter is a very rare complication. We experienced breakage of epidural catheter in patient of cancer pain during continuous epidural analgesia. We performed epidural catheterization between T12 and L1 level for pain control. After 6 weeks of epidural catheterization, abruptly patient complained severe pain. When we applied negative pressure to catheter by syringe before epidural injection of bolus dose, a few airs were aspirated. We strongly suspected breakage of catheter, and the breakage in 1617 cm from epidural catheter tip was confirmed by normal saline administration via epidural catheter. After new epidural catheter inserted again, patient's pain was controlled by VAS 24 and another complications were not appeared. We must carefully manage the catheter because of complications of epidural catheter.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Jeringas
/
Inyecciones Epidurales
/
Cateterismo
/
Analgesia Epidural
/
Catéteres
Límite:
Humanos
Idioma:
Coreano
Revista:
Korean Journal of Anesthesiology
Año:
2007
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS