Your browser doesn't support javascript.
loading
Assessment of Diffusion-Weighted Imaging-FLAIR Mismatch: Comparison between Conventional FLAIR versus Shorter-Repetition-Time FLAIR at 3T
Article en En | WPRIM | ID: wpr-194484
Biblioteca responsable: WPRO
ABSTRACT
PURPOSE: Fluid-attenuated inversion recovery (FLAIR) imaging can be obtained faster with shorter repletion time (TR), but it gets noisier. We hypothesized that shorter-TR FLAIR obtained at 3 tesla (3T) with a 32-channel coil may be comparable to conventional FLAIR. The aim of this study was to compare the diagnostic value between conventional FLAIR (TR = 9000 ms, FLAIR9000) and shorter-TR FLAIR (TR = 6000 ms, FLAIR6000) at 3T in terms of diffusion-weighted imaging-FLAIR mismatch. MATERIALS AND METHODS: We recruited 184 patients with acute ischemic stroke (28 patients 0.05). Interobserver agreement was excellent for both DWI-FLAIR9000 and DWI-FLAIR6000 (k = 0.904 and 0.883, respectively). Between the two FLAIR imaging sets, there was no significant difference of signal intensity ratio (mean, standard deviation; 1.25 ± 0.20; 1.24 ± 0.20, respectively) (P > 0.05). CONCLUSION: For the determination of mismatch or match between DWI and FLAIR imaging, there is no significant difference between FLAIR9000 and FLAIR6000 at 3T with a 32-channel coil.
Asunto(s)
Palabras clave
Texto completo: 1 Índice: WPRIM Asunto principal: Imagen por Resonancia Magnética / Accidente Cerebrovascular Límite: Humans Idioma: En Revista: Investigative Magnetic Resonance Imaging Año: 2016 Tipo del documento: Article
Texto completo: 1 Índice: WPRIM Asunto principal: Imagen por Resonancia Magnética / Accidente Cerebrovascular Límite: Humans Idioma: En Revista: Investigative Magnetic Resonance Imaging Año: 2016 Tipo del documento: Article