A Reinforced Endotracheal Tube Completely Severed by a Patient Bite and Lodged in the Right Main Bronchus / 대한중환자의학회지
Korean Journal of Critical Care Medicine ; : 70-73, 2017.
Artículo en Inglés | WPRIM | ID: wpr-194698
ABSTRACT
Reinforced endotracheal tubes (ETTs) are designed to resist kinking or compression. However, these have a potential risk of being obstructed or severed by a patient's bite. We report a case in which a reinforced ETT was severed by tube-bite while the patient was in the prone position during an intensive care unit stay. Bronchoscopic evaluation showed that the severed distal part of the tube had lodged in the patient's right main bronchus, and it had to be surgically removed. The patency of reinforced ETTs should be carefully monitored in patients intubated in the prone position.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Bronquios / Broncoscopía / Posición Prona / Derechos del Paciente / Obstrucción de las Vías Aéreas / Unidades de Cuidados Intensivos / Intubación Límite: Humanos Idioma: Inglés Revista: Korean Journal of Critical Care Medicine Año: 2017 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Bronquios / Broncoscopía / Posición Prona / Derechos del Paciente / Obstrucción de las Vías Aéreas / Unidades de Cuidados Intensivos / Intubación Límite: Humanos Idioma: Inglés Revista: Korean Journal of Critical Care Medicine Año: 2017 Tipo del documento: Artículo