Your browser doesn't support javascript.
loading
Performance Evaluation of the Autokit Total Ketone Bodies
Journal of Laboratory Medicine and Quality Assurance ; : 178-180, 2017.
Artículo en Coreano | WPRIM | ID: wpr-209174
ABSTRACT
In this study, the Autokit Total Ketone Bodies kit (Wako Pure Chemical, Japan), a total ketone measurement assay using an enzymatic method, was evaluated using a Roche Cobas e702 instrument (Roche Diagnostics, Germany). Precision, linearity, carryover, and reference range verification were evaluated with reference to Clinical Laboratory Standards Institute guidelines. Standard materials provided by the manufacturer and patient samples were used for the evaluation. The precision and carryover of the evaluation result satisfied the acceptance criteria. Linearity was also acceptable at more than 0.99. The quantitative Autokit Total Ketone Bodies kit is precise, and can be widely used in clinical laboratories.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Valores de Referencia / Ácido 3-Hidroxibutírico / Estudios de Evaluación como Asunto / Cuerpos Cetónicos / Métodos Tipo de estudio: Estudio diagnóstico / Estudios de evaluación / Guía de Práctica Clínica Límite: Humanos Idioma: Coreano Revista: Journal of Laboratory Medicine and Quality Assurance Año: 2017 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Valores de Referencia / Ácido 3-Hidroxibutírico / Estudios de Evaluación como Asunto / Cuerpos Cetónicos / Métodos Tipo de estudio: Estudio diagnóstico / Estudios de evaluación / Guía de Práctica Clínica Límite: Humanos Idioma: Coreano Revista: Journal of Laboratory Medicine and Quality Assurance Año: 2017 Tipo del documento: Artículo