Your browser doesn't support javascript.
loading
Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants
The Journal of Advanced Prosthodontics ; : 62-69, 2016.
Artículo en Inglés | WPRIM | ID: wpr-211465
ABSTRACT

PURPOSE:

The purpose of this study was to evaluate the effects of abutment screw length and cyclic loading on the removal torque (RTV) in external hex (EH) and internal hex (IH) implants. MATERIALS AND

METHODS:

Forty screw-retained single crowns were connected to external and internal hex implants. The prepared titanium abutment screws were classified into 8 groups based on the number of threads (n = 5 per group) EH 12.5, 6.5, 3.5, 2.5 and IH 6.5, 5, 3.5, 2.5 threads. The abutment screws were tightened with 20 Ncm torque twice with 10-minute intervals. After 5 minutes, the initial RTVs of the abutment screws were measured with a digital torque gauge (MGT12). A customized jig was constructed to apply a load along the implant long axis at the central fossa of the maxillary first molar. The post-loading RTVs were measured after 16,000 cycles of mechanical loading with 50 N at a 1-Hz frequency. Statistical analysis included one-way analysis of variance and paired t-tests.

RESULTS:

The post-loading RTVs were significantly lower than the initial RTVs in the EH 2.5 thread and IH 2.5 thread groups (P<.05). The initial RTVs exhibited no significant differences among the 8 groups, whereas the post-loading RTVs of the EH 6.5 and EH 3.5 thread groups were higher than those of the IH 3.5 thread group (P<.05).

CONCLUSION:

Within the limitations of this study, the external hex implants with short screw lengths were more advantageous than internal hex implants with short screw lengths in torque maintenance after cyclic loading.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Vértebra Cervical Axis / Titanio / Torque / Coronas / Diseño de Implante Dental-Pilar / Diente Molar Idioma: Inglés Revista: The Journal of Advanced Prosthodontics Año: 2016 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Vértebra Cervical Axis / Titanio / Torque / Coronas / Diseño de Implante Dental-Pilar / Diente Molar Idioma: Inglés Revista: The Journal of Advanced Prosthodontics Año: 2016 Tipo del documento: Artículo