Comparison of Dynamic Contour Tonometry and Goldmann Applanation Tonometry
Journal of the Korean Ophthalmological Society ; : 242-246, 2009.
Artículo en Coreano | WPRIM | ID: wpr-211853
ABSTRACT

PURPOSE:

To compare dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) and investigate the influence of central corneal thickness (CCT) onintraocular pressure.

METHODS:

In a prospective study of 165 eyes with glaucoma (135 eyes), glaucoma suspect (14 eyes), and ocular hypertension (16 eyes), intraocularpressure was measured with DCT and GAT, and followed by measurement of the CCT with ultrasound pachymetry. Statistical analysis were performed with simple linear regression analysis and t-test using SPSS (Statistical software, ver. 10; SPSS Inc., Chicago, IL).

RESULTS:

A clear correlation between DCT and GAT was found (r=0.733, p<0.001). The average intraocular pressure was 14.92+/-2.28 mmHg with DCT and 13.97+/-3.12 mmHg with GAT, and the intraocular pressure with DCT was 0.95+/-2.49 mmHg higher than with GAT. A meaningful correlation was shown between GAT and CCT (r2=0.145, p<0.001), but was not demonstrated between DCT and CCT (r2=0.012, p=0.081). In addition, the difference of intraocular pressure between GAT and DCT (GAT-DCT) showed a significant correlation with CCT (r2=0.145, p<0.001).

CONCLUSIONS:

DCT appears to be a reliable method for intraocular pressure measurement, which is not influenced by CCT, unlike GAT.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Péptidos / Modelos Lineales / Glaucoma / Hipertensión Ocular / Chicago / Estudios Prospectivos / Ojo / Presión Intraocular / Manometría Tipo de estudio: Estudio observacional País/Región como asunto: America del Norte Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Año: 2009 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Péptidos / Modelos Lineales / Glaucoma / Hipertensión Ocular / Chicago / Estudios Prospectivos / Ojo / Presión Intraocular / Manometría Tipo de estudio: Estudio observacional País/Región como asunto: America del Norte Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Año: 2009 Tipo del documento: Artículo