Your browser doesn't support javascript.
loading
Successful coronary stent retrieval from the ascending aorta using a gooseneck snare kit
The Korean Journal of Internal Medicine ; : 481-485, 2013.
Artículo en Inglés | WPRIM | ID: wpr-212575
ABSTRACT
Coronary stent dislodgement is a rare complication of percutaneous coronary intervention. We report a case of stent dislodgement in the ascending thoracic aorta. The stent was mechanically distorted in the left circumflex artery (LCX) while being delivered to the proximal LCX lesion. The balloon catheter was withdrawn, but the stent with the guide wire was remained in the ascending thoracic aorta. The stent was unable to be retrieved into the guide catheter, as it was distorted. A goose neck snare was used successfully to catch the stent in the ascending thoracic aorta and retrieved the stent externally via the arterial sheath.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Aorta Torácica / Cateterismo Cardíaco / Angioplastia Coronaria con Balón / Radiografía Intervencional / Stents / Resultado del Tratamiento / Angiografía Coronaria / Remoción de Dispositivos / Cuerpos Extraños Límite: Humanos / Masculino Idioma: Inglés Revista: The Korean Journal of Internal Medicine Año: 2013 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Aorta Torácica / Cateterismo Cardíaco / Angioplastia Coronaria con Balón / Radiografía Intervencional / Stents / Resultado del Tratamiento / Angiografía Coronaria / Remoción de Dispositivos / Cuerpos Extraños Límite: Humanos / Masculino Idioma: Inglés Revista: The Korean Journal of Internal Medicine Año: 2013 Tipo del documento: Artículo