Your browser doesn't support javascript.
loading
Fracture of J-tipped guidewire during central venous catheterization and its successful removal under fluoroscopic guidance: A case report / 대한마취과학회지
Korean Journal of Anesthesiology ; : 457-460, 2012.
Artículo en Inglés | WPRIM | ID: wpr-227534
ABSTRACT
Central venous catheterization by Seldinger's technique uses a guidewire which may cause complications such as kinking, knotting and fracture. Fractured guidewire may lead to severe outcomes such as embolization, and removal of it may also cause problems such as vessel damage. We experienced a case of right internal jugular venous catheterization complicated by guidewire fracture entrapped in the central venous catheter, and its successful removal under fluoroscopic guidance using snare-loop. The patient recovered without any complications. When resistance is felt during insertion or withdrawal of the guidewire, force should not be applied to the guidewire and care should also be exercised when passing the tissue dilator over the guidewire. Clinicians should be aware of this rare complication and snare-loop technique could be considered as one of the methods for removal of the fractured guidewire.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Cateterismo Venoso Central / Cateterismo / Catéteres / Catéteres Venosos Centrales / Glicosaminoglicanos Tipo de estudio: Guía de Práctica Clínica Límite: Humanos Idioma: Inglés Revista: Korean Journal of Anesthesiology Año: 2012 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Cateterismo Venoso Central / Cateterismo / Catéteres / Catéteres Venosos Centrales / Glicosaminoglicanos Tipo de estudio: Guía de Práctica Clínica Límite: Humanos Idioma: Inglés Revista: Korean Journal of Anesthesiology Año: 2012 Tipo del documento: Artículo