The measurement of radiation dosage in diagnostic x-ray device / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 1256-1259, 2007.
Artículo en Chino | WPRIM | ID: wpr-230707
ABSTRACT
In order to improve the diagnostic quality of X-ray device and to promote the implementation of radiation protection in accordance to the ICRU definition and ICRP classification standards for measuring radiation harm to human body, we have made a number of tests, including the kinetic energy released by ionizing air, the Half-value layer, the output dose coincidence, the High contrast resolution, the Consistency between visual light field and project X-ray field, and the Focus of X-ray tube. We have also analysed the measured data and results. The tests were conducted on the XEB150L-20 made by Shimadzu Co. in the middle of 1990s.
Asunto(s)
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Control de Calidad / Dosis de Radiación / Protección Radiológica / Radiometría / Radiografía / Monitoreo de Radiación / Métodos Tipo de estudio: Estudio diagnóstico / Guía de Práctica Clínica Idioma: Chino Revista: Journal of Biomedical Engineering Año: 2007 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Control de Calidad / Dosis de Radiación / Protección Radiológica / Radiometría / Radiografía / Monitoreo de Radiación / Métodos Tipo de estudio: Estudio diagnóstico / Guía de Práctica Clínica Idioma: Chino Revista: Journal of Biomedical Engineering Año: 2007 Tipo del documento: Artículo