The measurement of radiation dosage in diagnostic x-ray device / 生物医学工程学杂志
Journal of Biomedical Engineering
;
(6): 1256-1259, 2007.
Artículo
en Chino
| WPRIM
| ID: wpr-230707
ABSTRACT
In order to improve the diagnostic quality of X-ray device and to promote the implementation of radiation protection in accordance to the ICRU definition and ICRP classification standards for measuring radiation harm to human body, we have made a number of tests, including the kinetic energy released by ionizing air, the Half-value layer, the output dose coincidence, the High contrast resolution, the Consistency between visual light field and project X-ray field, and the Focus of X-ray tube. We have also analysed the measured data and results. The tests were conducted on the XEB150L-20 made by Shimadzu Co. in the middle of 1990s.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Control de Calidad
/
Dosis de Radiación
/
Protección Radiológica
/
Radiometría
/
Radiografía
/
Monitoreo de Radiación
/
Métodos
Tipo de estudio:
Estudio diagnóstico
/
Guía de Práctica Clínica
Idioma:
Chino
Revista:
Journal of Biomedical Engineering
Año:
2007