Your browser doesn't support javascript.
loading
Research on influence of different syringe types on quality control of syringe pump / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 1316-1324, 2014.
Artículo en Chino | WPRIM | ID: wpr-234409
ABSTRACT
To evaluate the influence of different brands of syringes on performance parameters and application quality of syringe pump, we carried out matching experiment of infusion/syringe pump analyzer Rigel Multi-Flo, made by ALK Co. in Sweden, for 3 different types of syringes to measure the flow rate, occlusion alarm pressure, occlusion time and bolus volume of 26 syringe pumps. Data of quality control were analyzed with statistical method. The results demonstrated that there were significant differences in the flow rate, occlusion alarm pressure, occlusion time and bolus volume (P<0. 01) between those of common syringes and original syringes. This study points out that the health care providers should select compatible syringes for syringe pumps, which provides evidence and guidance to assure the reasonable and safe application of syringe pumps in clinical practice.
Asunto(s)
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Control de Calidad / Jeringas / Bombas de Infusión Tipo de estudio: Guía de Práctica Clínica Idioma: Chino Revista: Journal of Biomedical Engineering Año: 2014 Tipo del documento: Artículo