Falsely Elevated Postvoid Residual Urine Volume in Uterine Myoma
Annals of Rehabilitation Medicine ; : 332-336, 2017.
Artículo en Inglés | WPRIM | ID: wpr-25596
ABSTRACT
Precise measurement of postvoid residual (PVR) urine volume is a key factor in assessing patients with voiding dysfunction, including those with lower urinary tract problems. The safe and noninvasive ultrasound bladder scan is the preferred mode to measure PVR volume. However, this procedure has a false-positive rate up to 9%, in the presence of ovarian cysts, renal cysts, ascites, or uterine myoma with cystic degeneration. Until now, cystic lesions are known to cause false positivity in ultrasound bladder scanner. However, we encountered falsely-elevated PVR in two cases of non-cystic uterine myomas. We present these cases with detailed radiologic images and volume measurement data.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Quistes Ováricos / Ascitis / Sistema Urinario / Vejiga Urinaria / Ultrasonografía / Leiomioma / Mioma Tipo de estudio: Estudio diagnóstico Límite: Femenino / Humanos Idioma: Inglés Revista: Annals of Rehabilitation Medicine Año: 2017 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Quistes Ováricos / Ascitis / Sistema Urinario / Vejiga Urinaria / Ultrasonografía / Leiomioma / Mioma Tipo de estudio: Estudio diagnóstico Límite: Femenino / Humanos Idioma: Inglés Revista: Annals of Rehabilitation Medicine Año: 2017 Tipo del documento: Artículo