The research on high-density flexible microelectrode array of retinal prosthesis based on MEMS technology / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 407-410, 2013.
Artículo en Chino | WPRIM | ID: wpr-264175
ABSTRACT
The paper proposed a new method to design and fabricate a flexible neural microelectrode arrays (MEA) for retinal prosthesis, the ion-beam technology was introduced to decrease the width of conductive wires and the distances between wires, a high density MEA (120 microelectrodes with a matrix of 10 x 12) was fabricated on a single layer of polymer from this. The MEA was proved to possess a fine electrochemical property. In vitro test, the average impedance of MEA on 1 kHz was 16 k omega +/- 2 k omega and the average phase difference was -85 degrees +/-30 degrees.
Asunto(s)
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Diseño de Prótesis / Sistemas Microelectromecánicos / Prótesis Visuales / Microelectrodos Idioma: Chino Revista: Chinese Journal of Medical Instrumentation Año: 2013 Tipo del documento: Artículo