Your browser doesn't support javascript.
loading
Establishment and evaluation of intraocular lens loop support force test platform / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 291-293, 2013.
Artículo en Chino | WPRIM | ID: wpr-264209
ABSTRACT
<p><b>OBJECTIVE</b>Establish the test platform of the intraocular lens loop, and the platform was evaluated through the experiment.</p><p><b>METHODS</b>The intraocular lens loop test platform is made up with three models. The different intraocular lens haptics support force can be completed by replacing different sample holder model.</p><p><b>RESULTS</b>The standard deviation and the coefficient of variation were calculated through the result of the fifteen samples. The standard deviation was 0.04 mN, and the coefficient of variation was 0.66%. The two values were in the acceptable range.</p><p><b>CONCLUSIONS</b>The platform was so stabilizing that it could be used to test support force of IOL loop. The different shapes of IOL could be tested on the platform through the replacement of the holder model.</p>
Asunto(s)
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Diseño de Prótesis / Lentes Intraoculares Idioma: Chino Revista: Chinese Journal of Medical Instrumentation Año: 2013 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Diseño de Prótesis / Lentes Intraoculares Idioma: Chino Revista: Chinese Journal of Medical Instrumentation Año: 2013 Tipo del documento: Artículo