Study on Accurately Controlling Discharge Energy Method Used in External Defibrillator / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 17-21, 2016.
Artículo
en Chino
| WPRIM
| ID: wpr-265583
ABSTRACT
This paper introduces a new method which controls discharge energy accurately. It is achieved by calculating target voltage based on transthoracic impedance and accurately controlling charging voltage and discharge pulse width. A new defibrillator is designed and programmed using this method. The test results show that this method is valid and applicable to all kinds of external defibrillators.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Estándares de Referencia
/
Desfibriladores
/
Diseño de Equipo
Idioma:
Chino
Revista:
Chinese Journal of Medical Instrumentation
Año:
2016
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS