Your browser doesn't support javascript.
loading
Study on Accurately Controlling Discharge Energy Method Used in External Defibrillator / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 17-21, 2016.
Artículo en Chino | WPRIM | ID: wpr-265583
ABSTRACT
This paper introduces a new method which controls discharge energy accurately. It is achieved by calculating target voltage based on transthoracic impedance and accurately controlling charging voltage and discharge pulse width. A new defibrillator is designed and programmed using this method. The test results show that this method is valid and applicable to all kinds of external defibrillators.
Asunto(s)
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Estándares de Referencia / Desfibriladores / Diseño de Equipo Idioma: Chino Revista: Chinese Journal of Medical Instrumentation Año: 2016 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS